FDA Adverse Event Injury Summary report: N

ZEPHIR ANTERIOR CERVICAL SYSTEM

MDR report key: 7205035 · Received January 19, 2018

Report

Report Number
1030489-2018-00087
Event Type
Injury
Date Received
January 19, 2018
Date of Event
August 14, 2017
Report Date
January 19, 2018
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
KWP
PMA / PMN Number
SEE H10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS PART IS NOT APPROVED FOR USE IN THE US, HOWEVER A LIKE DEVICE WITH PART# 8792113, 510K# K994239 AND (B)(4) IS CLEARED FOR US. PRODUCT ANALYSIS: THE PLATE AND FOUR BONE SCREWS WHERE RETURNED. THE BOTH LOCKING TABS HAVE BEEN BROKEN OFF ONE SIDE OF THE PLATE. ON ONE OF THE BONES SCREWS (LOT# H5174732) HAS CLEAR WITNESS MARKS WHERE IT IS PUSHED UPWARDS AGAINST THE TAB AND BROKE IT. ANOTHER SCREW ALSO LOT# H5174732 HAS WITNESS MARKS AROUND THE MINOR DIAMETER OF THE SCREW INDICATING THAT SCREW MAY NOT HAVE BEEN AT THE PERFECT ANGLE DURING IMPLANTATION. THERE DOES NOT APPEAR TO BE A MANUFACTURING ISSUE WITH THE SCREWS OR THE PLATE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

PROCEDURE USED IN INITIAL SURGERY: C5/6: ANTERIOR CERVICAL DECOMPRESSION FUSION PRE-OP DIAGNOSIS FOR REVISION: SCREW BACK OUT PROCEDURE USED IN REVISION: ANTERIOR CERVICAL DECOMPRESSION FUSION, C5/6 IT WAS REPORTED THAT THE PATIENT UNDERWENT A SURGERY AT C5/6 IN WHICH THE PLATE WAS IMPLANTED. POST-OP, AFTER APPROXIMATELY SIX MONTHS SCREW BACKED-OUT. PATIENT UNDERWENT REVISION SURGERY FOR REPLACEMENT OF THE SCREW. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE REVISION SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48495 ZEPHIR ANTERIOR CERVICAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MDT SOFAMOR DANEK PUERTO RICO MFG NA H5174732

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention