PRODUCT COMPLAINT#: (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. INVESTIGATION SUMMARY: THE COMPLAINT DEVICE IS NOT AVAILABLE FOR PHYSICAL EVALUATION, HENCE, THE COMPLAINT CANNOT BE CONFIRMED. A MEMBER OF NEW PRODUCT DEVELOPMENT (NPD) WAS CONSULTED FOR FURTHER INSIGHT INTO THE FAILURE MODE. THE NPD ENGINEER INDICATED THAT SOME POTENTIAL CAUSES FOR THE ANCHOR BEING DIFFICULT TO DEPLOY COULD BE HARDER THAN USUAL PATIENT BONE, OR THE SURGEON NOT DRILLING A BIG ENOUGH HOLE FOR THE ANCHOR THUS CAUSING A TIGHTER FIT THAN NORMAL, OR THE SURGEON THOUGHT THEY USED THE CORRECT DRILL SIZE BUT IT WAS SMALLER. HOWEVER, GIVEN THE INFORMATION PROVIDED WE CANNOT DISCERN A DEFINITIVE ROOT CAUSE FOR THE REPORTED FAILURE. A NON-CONFORMANCE SEARCH WAS CONDUCTED TO INVESTIGATE ANY DEFECTS DURING PRODUCTION IDENTIFIED THAT MAY CONTRIBUTE TO THE COMPLAINT CONDITION. NO NON-CONFORMANCE WAS IDENTIFIED FOR THIS PART-LOT NUMBER COMBINATION. AT THIS POINT, NO CORRECTIVE ACTION IS REQUIRED AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. DEVICE HISTORY LOT: NO NON-CONFORMANCE WAS IDENTIFIED FOR THIS PART/LOT NUMBER COMBINATION PER QLIK QUERY EXECUTED 11/01/2018. REVIEW CONDUCTED PER NR-0108348. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UDI: (B)(4).
PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UDI:(B)(4).
ADDITIONAL INFORMATION RECEIVED ON 11/1/2018. THE SALES REP COMMUNICATED THAT THE 3 BRAND NEW DEVICES WERE THE REMAINDER OF THE LOT, WHICH WERE REQUESTED BY THE CUSTOMER TO BE SENT BACK. HOWEVER, THE COMPLAINT DEVICES WERE DISCARDED.
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.¿ (B)(4). THE EXPIRATION DATE IS NOT CURRENTLY AVAILABLE. ASSOCIATED MEDWATCHES: 1221934-2018-50072 AND 1221934-2018-50074.
PRODUCT COMPLAINT # (B)(4). THREE BRAND NEW UNUSED DEVICES WERE RECEIVED FROM THE SALES REPRESENTATIVE. HE WAS CONTACTED TO RETURNED THE COMPLAINT DEVICES FOR EVALUATION. UDI: (B)(4). THE EXPIRATION DATE IS NOT CURRENTLY AVAILABLE. ASSOCIATED MEDWATCHES: 1221934-2018-50072 AND 1221934-2018-50074.
IT WAS REPORTED BY THE CUSTOMER THAT THREE OF THERE MINI QUICK ANCHOR #2/0 ORTHOCORD V-5 WERE DEPLOYING TOO HARD DURING AN UNKNOWN PROCEDURE. THE CASE WAS COMPLETED WITH ANOTHER LIKE DEVICE. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS. THE DEVICES ARE BEING RETURNED.
IT WAS REPORTED BY THE CUSTOMER THAT THREE OF THERE MINI QUICK ANCHOR #2/0 ORTHOCORD V-5 WERE DEPLOYING TOO HARD DURING AN UNKNOWN PROCEDURE. THE CASE WAS COMPLETED WITH ANOTHER LIKE DEVICE. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS. THE DEVICES ARE BEING RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 43101 | MINI QA+ #2/O OCORD V-5 | SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE | JDR | DEPUY MITEK LLC US | L302230 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|