NEURON MAX 6F 088 LONG SHEATH
Report
- Report Number
- 3005168196-2018-00147
- Event Type
- Injury
- Date Received
- January 16, 2018
- Date of Event
- December 8, 2017
- Report Date
- December 21, 2017
- Manufacturer
- PENUMBRA, INC.
- Product Code
- DQY
- PMA / PMN Number
- K111380
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- PHYSICIAN
Narratives
THIS DEVICE IS AVAILABLE FOR RETURN. A FOLLOW UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE INVESTIGATION.
DATE OF EVENT: TO (B)(6) 2017. RESULTS: THE RETURNED PIECE WAS A 3MM PIECE OF EXTRUSION FROM THE NEURON MAX. CONCLUSIONS: THE FINAL CONCLUSIONS OF THE DEVICE INVESTIGATION ARE AS FOLLOWS: EVALUATION OF THE RETURNED FRAGMENT OF THE DEVICE REVEALED THAT A PIECE OF THE NEURON MAX WAS FRACTURED FROM THE CATHETER SHAFT. INSPECTION OF THE FRACTURE REVEALED THAT IT DID NOT OCCUR AT A MATERIAL JUNCTION. FURTHER EVALUATION OF THE FRACTURED DISTAL PIECE SHOWS THAT THE EXTRUSION WAS PROPERLY LAMINATED OVER THE MARKER BAND. THE ROOT CAUSE OF THIS DAMAGE COULD NOT BE DETERMINED. PENUMBRA CATHETERS ARE VISUALLY INSPECTED DURING IN-PROCESS INSPECTION AND DURING QUALITY INSPECTION AFTER MANUFACTURING. THE PRODUCT LOT NUMBER WAS NOT PROVIDED, THEREFORE, THE MANUFACTURING RECORDS COULD NOT BE REVIEWED.
THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A NEURON MAX. DURING THE PROCEDURE, A PIECE BROKE OFF OF THE NEURON MAX AND MIGRATED, OCCLUDING THE MCA. THE PATIENT PASSED AWAY A FEW DAYS LATER IN THE HOSPITAL.
THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX). DURING THE PROCEDURE, WHILE USING A NON-PENUMBRA CATHETER WITHIN THE NEURON MAX, A PIECE OF THE TIP OF THE NEURON MAX BROKE OFF AND MIGRATED, OCCLUDING THE MCA; HOWEVER, THE BROKEN TIP OF THE NEURON MAX WAS NOT NOTICED UNTIL AN X-RAY WAS PERFORMED THE NEXT DAY. THE PATIENT PASSED AWAY IN THE HOSPITAL FROM BRONCHOPNEUMONIA A FEW DAYS LATER. THE EXACT DATE THAT THE PROCEDURE TOOK PLACE, DURING WHICH THE DISTAL TIP OF THE NEURON MAX BROKE OFF AND MIGRATED, IS UNKNOWN; HOWEVER, THE PHYSICIAN STATED THAT THE PATIENT DIED OF BRONCHOPNEUMONIA ON (B)(6) 2017, AFTER A FEW DAYS IN THE HOSPITAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 40539 | NEURON MAX 6F 088 LONG SHEATH | DQY | DQY | PENUMBRA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death| L |