FDA Adverse Event Injury Summary report: N

NEURON MAX 6F 088 LONG SHEATH

MDR report key: 7197873 · Received January 16, 2018

Report

Report Number
3005168196-2018-00147
Event Type
Injury
Date Received
January 16, 2018
Date of Event
December 8, 2017
Report Date
December 21, 2017
Manufacturer
PENUMBRA, INC.
Product Code
DQY
PMA / PMN Number
K111380
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS AVAILABLE FOR RETURN. A FOLLOW UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE INVESTIGATION.

Additional Manufacturer Narrative · 1

DATE OF EVENT: TO (B)(6) 2017. RESULTS: THE RETURNED PIECE WAS A 3MM PIECE OF EXTRUSION FROM THE NEURON MAX. CONCLUSIONS: THE FINAL CONCLUSIONS OF THE DEVICE INVESTIGATION ARE AS FOLLOWS: EVALUATION OF THE RETURNED FRAGMENT OF THE DEVICE REVEALED THAT A PIECE OF THE NEURON MAX WAS FRACTURED FROM THE CATHETER SHAFT. INSPECTION OF THE FRACTURE REVEALED THAT IT DID NOT OCCUR AT A MATERIAL JUNCTION. FURTHER EVALUATION OF THE FRACTURED DISTAL PIECE SHOWS THAT THE EXTRUSION WAS PROPERLY LAMINATED OVER THE MARKER BAND. THE ROOT CAUSE OF THIS DAMAGE COULD NOT BE DETERMINED. PENUMBRA CATHETERS ARE VISUALLY INSPECTED DURING IN-PROCESS INSPECTION AND DURING QUALITY INSPECTION AFTER MANUFACTURING. THE PRODUCT LOT NUMBER WAS NOT PROVIDED, THEREFORE, THE MANUFACTURING RECORDS COULD NOT BE REVIEWED.

Description of Event or Problem · 1

THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A NEURON MAX. DURING THE PROCEDURE, A PIECE BROKE OFF OF THE NEURON MAX AND MIGRATED, OCCLUDING THE MCA. THE PATIENT PASSED AWAY A FEW DAYS LATER IN THE HOSPITAL.

Description of Event or Problem · 1

THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX). DURING THE PROCEDURE, WHILE USING A NON-PENUMBRA CATHETER WITHIN THE NEURON MAX, A PIECE OF THE TIP OF THE NEURON MAX BROKE OFF AND MIGRATED, OCCLUDING THE MCA; HOWEVER, THE BROKEN TIP OF THE NEURON MAX WAS NOT NOTICED UNTIL AN X-RAY WAS PERFORMED THE NEXT DAY. THE PATIENT PASSED AWAY IN THE HOSPITAL FROM BRONCHOPNEUMONIA A FEW DAYS LATER. THE EXACT DATE THAT THE PROCEDURE TOOK PLACE, DURING WHICH THE DISTAL TIP OF THE NEURON MAX BROKE OFF AND MIGRATED, IS UNKNOWN; HOWEVER, THE PHYSICIAN STATED THAT THE PATIENT DIED OF BRONCHOPNEUMONIA ON (B)(6) 2017, AFTER A FEW DAYS IN THE HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40539 NEURON MAX 6F 088 LONG SHEATH DQY DQY PENUMBRA, INC.

Patients

Seq Age Sex Outcome Treatment
1 Death| L