BD SAF-T-INTIMA IV CATHETER SAFETY SYSTEM
Report
- Report Number
- 9610847-2017-00232
- Event Type
- Injury
- Date Received
- January 16, 2018
- Date of Event
- December 26, 2017
- Report Date
- February 19, 2018
- Manufacturer
- BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
- Product Code
- FOZ
- PMA / PMN Number
- K923702
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
DATE OF EVENT: UNKNOWN. INITIAL REPORTER'S PHONE#: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION RESULTS: ONE REPRESENTATIVE UNIT WAS RECEIVED FOR ANALYSIS. OUR QUALITY ENGINEER VISUALLY INSPECTED THE RETURNED UNIT AND NO PHYSICAL OR MECHANICAL DAMAGE OR FOREIGN MATTER WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF REPORTED LOT NUMBER 6243989. WE WERE UNABLE TO DETERMINE A ROOT CAUSE FOR THIS EVENT. REPORTED DEFECT WAS NOT CONFIRMED IN REPRESENTATIVE SAMPLE; SINCE, DURING THE INSPECTION OF THIS SAMPLE WE DO NOT FIND FOREIGN MATTER IN THE CATHETER TIP OR NEEDLE TIP. CERTIFICATE OF STERILITY FOR THIS REPORTED LOT NUMBER 6243989 WAS REVIEWED AND ALL THE UNITS (B)(4) MEET WITH ACCEPTANCE CRITERIA. WE CANNOT CONFIRM OR ASSOCIATE THIS REPORTED DEFECT TO MANUFACTURING PROCESS. NO CAPA WAS OPENED SINCE THIS ISSUE COULD NOT BE CONFIRMED AS MANUFACTURING RELATED.
IT WAS REPORTED AFTER PLACEMENT OF THE BD SAF-T-INTIMA IV CATHETER SAFETY SYSTEM, ON SEVERAL OCCASIONS, ¿ALLERGIC REACTION, ABSCESS, INFECTION¿ WAS NOTED. THERE WAS NO REPORT OF ANY MEDICAL INTERVENTIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 36788 | BD SAF-T-INTIMA IV CATHETER SAFETY SYSTEM | INTRAVASCULAR CATHETER | FOZ | BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. | 6243989 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |