Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. INITIAL REPORTER'S PHONE#: (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

INVESTIGATION RESULTS: ONE REPRESENTATIVE UNIT WAS RECEIVED FOR ANALYSIS. OUR QUALITY ENGINEER VISUALLY INSPECTED THE RETURNED UNIT AND NO PHYSICAL OR MECHANICAL DAMAGE OR FOREIGN MATTER WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND REVEALED NO IRREGULARITIES DURING THE MANUFACTURE OF REPORTED LOT NUMBER 6243989. WE WERE UNABLE TO DETERMINE A ROOT CAUSE FOR THIS EVENT. REPORTED DEFECT WAS NOT CONFIRMED IN REPRESENTATIVE SAMPLE; SINCE, DURING THE INSPECTION OF THIS SAMPLE WE DO NOT FIND FOREIGN MATTER IN THE CATHETER TIP OR NEEDLE TIP. CERTIFICATE OF STERILITY FOR THIS REPORTED LOT NUMBER 6243989 WAS REVIEWED AND ALL THE UNITS (B)(4) MEET WITH ACCEPTANCE CRITERIA. WE CANNOT CONFIRM OR ASSOCIATE THIS REPORTED DEFECT TO MANUFACTURING PROCESS. NO CAPA WAS OPENED SINCE THIS ISSUE COULD NOT BE CONFIRMED AS MANUFACTURING RELATED.

Description of Event or Problem · 1

IT WAS REPORTED AFTER PLACEMENT OF THE BD SAF-T-INTIMA IV CATHETER SAFETY SYSTEM, ON SEVERAL OCCASIONS, ¿ALLERGIC REACTION, ABSCESS, INFECTION¿ WAS NOTED. THERE WAS NO REPORT OF ANY MEDICAL INTERVENTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
36788 BD SAF-T-INTIMA IV CATHETER SAFETY SYSTEM INTRAVASCULAR CATHETER FOZ BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 6243989

Patients

Seq Age Sex Outcome Treatment
1 Other