FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 7195674 · Received January 16, 2018

Report

Report Number
3004753838-2018-005655
Event Type
Malfunction
Date Received
January 16, 2018
Date of Event
December 9, 2017
Report Date
December 21, 2017
Manufacturer
DEXCOM, INC.
Product Code
PQM
UDI-DI
00386270000002
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PATIENT EXPERIENCED ADVERSE EVENTS ON DIFFERENT DAYS WITH THE SAME DEVICE. THE FOLLOWING REPORTS ARE BEING SUBMITTED: (B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT ON (B)(6) 2017, A LOSS OF CONNECTION BETWEEN THE TRANSMITTER AND SMART DEVICE OCCURRED. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. DATA WAS RETURNED AND EVALUATED. THE REPORTED EVENT OF A LOSS OF CONNECTION WAS CONFIRMED VIA DATA. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
40907 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR PQM DEXCOM, INC. N/A 00386270000002

Patients

Seq Age Sex Outcome Treatment
1 21 YR