UNIFY ASSURA CRT-D RF HV
Report
- Report Number
- 2017865-2018-00633
- Event Type
- Injury
- Date Received
- January 12, 2018
- Date of Event
- December 19, 2017
- Report Date
- May 17, 2018
- Manufacturer
- ST. JUDE MEDICAL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P970013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE IS INCLUDED IN THE PREMATURE BATTERY DEPLETION WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATOR ADVISORY NOTICE ISSUED BY ST. JUDE MEDICAL ON 11 OCTOBER 2016.
INTERROGATION OF THE DEVICE REVEALED IT WAS ABOVE ERI WHEN RECEIVED. BASED ON THIS INFORMATION, THE DEVICE WAS FOUND TO COMMUNICATE APPROPRIATELY WITH A MERLIN PROGRAMMER AND HAS NOT REACHED THE ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE IS INCLUDED IN THE PREMATURE BATTERY DEPLETION WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATOR ADVISORY NOTICE ISSUED BY ST. JUDE MEDICAL ON 11 OCTOBER 2016.
FOLLOWING THE ADVISORY FOR PREMATURE BATTERY DEPLETION WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATOR, ALTHOUGH THERE WAS NO ERI ALERT OR ALLEGATION OF PREMATURE BATTERY DEPLETION, THE DEVICE WAS EXPLANTED PROPHYLACTICALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 29518 | UNIFY ASSURA CRT-D RF HV | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC. | CD3357-40C | 4192704 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R |