FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 7185989
·
Received January 12, 2018
Report
- Report Number
- 3008642652-2018-00442
- Event Type
- Malfunction
- Date Received
- January 12, 2018
- Date of Event
- December 5, 2017
- Report Date
- January 10, 2018
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005036
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF BATTERY (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY FAULTS, WON'T POWER ON MONITOR) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY WAS UNABLE TO POWER ON A MONITOR OR CHARGE. THE CAUSE FOR THE FAILURE WAS ISOLATED OT A DEPLETED TRI-CELL (0.012V). THE ROOT CAUSE OF THE DEFECTIVE TRI-CELL WAS UNABLE TO BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK.
Description of Event or Problem · 1
A US DISTRIBUTOR REPORTED THAT A PATIENT WAS RECEIVING BATTERY FAULTS AND ONE OF THE PATIENT'S BATTERIES WOULDN'T POWER ON A MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 29645 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | WCD 4000 | 00855778005036 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |