PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 7182505 · Received January 11, 2018

Report

Report Number
2024168-2018-00289
Event Type
Injury
Date Received
January 11, 2018
Date of Event
December 26, 2017
Report Date
April 19, 2018
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). LOT # WAS CORRECTED FROM 7021541 TO 7021642. ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED FAILURE SUTURE DETACHMENT WAS CONFIRMED AS A LINK DETACHMENT/SUTURE RETRIEVAL ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEARS TO BE RELATED TO PROCEDURE CIRCUMSTANCE. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING OF THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE VIA A 6FR SHEATH, AFTER A DIAGNOSTIC PROCEDURE. REPORTEDLY, A SUTURE BREAK OCCURRED. A SECOND PROGLIDE DEVICE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27115 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 7021642

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention SHEATH: 6FR