FDA Adverse Event Malfunction Summary report: N

GUNTHER TULIP JUGULAR VENA CAVA FILTER SET

MDR report key: 7182298 · Received January 11, 2018

Report

Report Number
1820334-2017-04498
Event Type
Malfunction
Date Received
January 11, 2018
Report Date
March 28, 2018
Manufacturer
COOK INC
Product Code
DTK
UDI-DI
10827002330174
PMA / PMN Number
K072240
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

WILLIAM COOK (B)(4) INITIALLY REPORTED EVENT UNDER MANUFACTURER REPORT #3002808486-2017-02329 . NEW INFORMATION WAS RECEIVED IDENTIFYING THAT THE PRODUCT WAS A COOK INC. (B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION.

Additional Manufacturer Narrative · 1

INVESTIGATION - COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. A FILTER THAT IS EMBEDDED IN THE WALL OF THE IVC MAY BE DIFFICULT TO RETRIEVE. FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. FILTER TILT IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE AND MAY OCCUR DURING PLACEMENT OR DURING IMPLANTING PERIOD. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 1

PT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6) 2011 VIA THE RIGHT INTERNAL JUGULAR VEIN DUE TO A HISTORY OF DEEP VEIN THROMBOSIS AND PULMONARY EMBOLISM. PT IS ALLEGING DEVICE UNABLE TO BE RETRIEVED. DEVICE RETRIEVAL WAS ATTEMPTED ON (B)(6) 2012 AND (B)(6) 2012. THE RETRIEVAL ATTEMPTS WERE UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27074 GUNTHER TULIP JUGULAR VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK COOK INC G33017 10827002330174

Patients

Seq Age Sex Outcome Treatment
1