FDA Adverse Event Malfunction Summary report: N

BD¿ ALLERGY SYRINGE WITH NEEDLE, TRAY

MDR report key: 7180516 · Received January 10, 2018

Report

Report Number
1920898-2017-00430
Event Type
Malfunction
Date Received
January 10, 2018
Date of Event
December 11, 2017
Report Date
March 12, 2018
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
30382903055365
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: CUSTOMER RETURNED A PHOTO OF A TRAY OF 1/2ML, 10MM, 27G ALLERGY SYRINGES FROM LOT # 7072704. CUSTOMER STATES THAT THE LABEL HAD "TREATMENT" CROSSED OUT AND "TESTING" WAS WRITTEN. THE PHOTO WAS EXAMINED AND THE TRAY COVER EXHIBITED ¿TREATMENT¿ PRINTED ON IT. REFER TO CAPA (B)(4) AND SITUATION ANALYSIS (B)(4). BASED ON THE SAMPLES / PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: CONFIRMED: BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION IS REQUIRED AT THIS TIME. INVESTIGATION CONCLUSION: REFER TO CAPA (B)(4) AND SITUATION ANALYSIS (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LABEL OF A BD ALLERGY TRAY PRECISIONGLIDE¿ ATTACHED NEEDLE WITHOUT SAFETY WAS INCORRECT, AND STATED ¿TREATMENT¿ INSTEAD OF ¿TESTING¿ ON THE LABEL. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
24179 BD¿ ALLERGY SYRINGE WITH NEEDLE, TRAY SYRINGE WITH NEEDLE FMF BD MEDICAL - DIABETES CARE 7072704 30382903055365

Patients

Seq Age Sex Outcome Treatment
1 Other