FDA Adverse Event Malfunction Summary report: N

WATERPIK COMPLETE CARE

MDR report key: 7179261 · Received January 10, 2018

Report

Report Number
0001712259-2017-00005
Event Type
Malfunction
Date Received
January 10, 2018
Date of Event
December 18, 2017
Report Date
February 2, 2018
Manufacturer
WATER PIK, INC.
Product Code
EFS
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

PRODUCT EXPERIENCED ELECTRICAL FAILURE AND INTERNAL ARCING/BURNING, ALL DAMAGE WAS CONTAINED INSIDE THE UNIT. FAILURE ANALYSIS SHOWED THAT THE MOV COMPONENT FAILED AND CAUSED THE ELECTRICAL FAILURE. MOV'S ARE DESIGNED TO PROTECT THE CIRCUITRY BUT CAN EITHER BE OVERWHELMED OR WORN OUT AND RESULT IN THE TYPE OF FAILURE SEEN.

Description of Event or Problem · 0

PER THE CUSTOMER....CLAIMS THE DEVICE WAS PLUGGED IN BUT NOT IN USE WHEN UNIT BEGAN TO PRODUCE A BURNING SMELL. THE CUSTOMER CLAIMS THE PRODUCT PRODUCED A BLACK FILM RESIDUE ON THE COUNTERTOP AND MELTED THEIR WALL OUTLET. CUSTOMER IS SEEKING A REPLACEMENT PRODUCT OR REFUND.

Additional Manufacturer Narrative · 1

UNIT IN QUESTION HAS NOT BEEN RETURNED FOR TESTING AND EVALUATION.

Description of Event or Problem · 1

PER THE CUSTOMER....CLAIMS THE DEVICE WAS PLUGGED IN BUT NOT IN USE WHEN UNIT BEGAN TO PRODUCE A BURNING SMELL. THE CUSTOMER CLAIMS THE PRODUCT PRODUCED A BLACK FILM RESIDUE ON THE COUNTERTOP AND MELTED THEIR WALL OUTLET. CUSTOMER IS SEEKING A REPLACEMENT PRODUCT OR REFUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
24769 WATERPIK COMPLETE CARE ORAL IRRIGATION UNIT EFS WATER PIK, INC. WP-871

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other