FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 717863 · Received September 1, 2005

Report

Report Number
2243072-2005-00330
Event Type
Other
Date Received
September 1, 2005
Date of Event
August 4, 2005
Report Date
August 31, 2005
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT GETTING HIGHER THAN NORMAL READINGS ON HIS FAMILY MEMBER'S METER. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (226) AS REFERENCE POINT VS. BD READINGS (113, 338) YIELDED SOME DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 4064315

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other