FDA Adverse Event Malfunction Summary report: N

HYF HND SWTCH PENCL AUTOCLAVAB

MDR report key: 7175842 · Received January 9, 2018

Report

Report Number
1320894-2018-00002
Event Type
Malfunction
Date Received
January 9, 2018
Date of Event
December 15, 2017
Report Date
January 9, 2018
Manufacturer
CONMED CORPORATION
Product Code
GEI
PMA / PMN Number
K001159
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED DEVICE WAS DISCARDED BY THE USER FACILITY DUE TO BIOMATERIAL ON THE DEVICE; THEREFORE, AN EVALUATION IS UNABLE TO BE PERFORMED. A REVIEW OF THE MANUFACTURING DOCUMENTS WAS UNABLE TO BE PERFORMED AS THE LOT NUMBER WAS NOT PROVIDED. A TWO-YEAR HISTORICAL COMPLAINT REVIEW REVEALED ONE PRIOR COMPLAINT RELATED TO THIS FAILURE FOR THIS PRODUCT. IN THAT SAME TIMEFRAME, (B)(4) DEVICES HAVE BEEN MANUFACTURED AND SHIPPED WORLDWIDE. THE CLINICAL DATA REPORT (CDR) STATES THAT WHEN USED UNDER THE CONDITIONS AND FOR THE PURPOSES INTENDED THAT THE UNDESIRABLE SIDE EFFECTS, UNDER NORMAL CONDITIONS OF USE, ARE ACCEPTABLE WHEN WEIGHED AGAINST THE BENEFITS TO THE PATIENTS. THE INSTRUCTIONS FOR USE ADVISE THE USER OF THE FOLLOWING. -DO NOT USE THE DEVICE IF ANY SIGNS OF CRACKS, CUTS, PUNCTURES, NICKS OR ABRASION ARE VISIBLE. -PROLONGED USE OF THE HANDPIECE CAUSES THE ELECTRODE TO BECOME VERY HOT AND THE TIP REMAINS HOT FOR A PERIOD AFTER USE. -DO NOT ALLOW THE TIP TO COME IN CONTACT WITH THE PATIENT OR OTHERS AFTER ACTIVATION TO AVOID BURNS. THIS INCIDENT TYPE WILL CONTINUE TO BE MONITORED THROUGH THE COMPLAINT SYSTEM TO ASSURE PATIENT SAFETY.

Description of Event or Problem · 1

A CUSTOMER REPORTED DURING A 7-900-5 PENCIL "SPARKED AND CAUGHT FIRE AT THE PIN CONNECTOR". THIS INCIDENT OCCURRED DURING PRE-OP TESTING; THEREFORE, NO PATIENT CONTACT. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
17886 HYF HND SWTCH PENCL AUTOCLAVAB HYFRECATOR ACCESSORIES GEI CONMED CORPORATION UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1