FDA Adverse Event Malfunction Summary report: N

CATGUT PLAIN 3/0 (3) 75CM HR26

MDR report key: 7166043 · Received January 4, 2018

Report

Report Number
3003639970-2018-00009
Event Type
Malfunction
Date Received
January 4, 2018
Date of Event
November 23, 2017
Report Date
January 12, 2018
Manufacturer
B.BRAUN SURGICAL SA
Product Code
GAL
PMA / PMN Number
K991223
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

AESCULAP INC. (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF B. BRAUN SURGICAL S.A. (MANUFACTURER). EXEMPTION NUMBER: E2014012 MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING H3 OTHER TEXT : DEVICE NOT RETURNED

Additional Manufacturer Narrative · 0

SAMPLES RECEIVED: NO SAMPLES ARE RECEIVED FOR THE INVESTIGATION; ATTACHED IMAGE OF THE POSTERIOR PACKAGING OF THE REPORTED SUTURE IS ATTACHED. PRELIMINARY ANALYSIS: · STOCK REVIEW: WE PROCEED TO REVIEW THE AVAILABLE UNITS OF THIS PRODUCT CODE AND LOT NUMBER, EVIDENCING THAT TO DATE THERE ARE NO AVAILABLE UNITS IN OUR STOCK. · PRODUCED / IMPORTED QUANTITY: FROM THIS PRODUCT CODE AND BATCH NUMBER, (B)(4) UNITS IN TOTAL WERE MANUFACTURED IN BBMCO. · BACKGROUND: PROCEED TO REVIEW THE DATABASE OF CLAIMS AND VERIFY THAT TO DATE THERE ARE NO REPORTS OF INCIDENTS, RELATED TO THIS PRODUCT CODE LOT NUMBER AND CAUSE. · BATCH RECORD / BATCH RECORD: A REVIEW OF THE DOCUMENTATION CORRESPONDING TO THE MANUFACTURING OF THE BATCH WAS MADE, IN WHICH THE CONTROL IN PROCESS AND CONTROL OF THE FINISHED PRODUCT WERE VERIFIED. NO DEVIATIONS OR ALTERATIONS ARE IDENTIFIED DURING THE PROCESS. ANALYSIS OF THE SAMPLES: THE RECEIVED IMAGE CORRESPONDS TO THE REPORTED LOT; IT WAS VERIFIED VISUALLY AND IT IS EVIDENT THAT THE OUTER PACKAGING PRESENTS SPOTS, CAUSED BY THE LEAKAGE OF THE STERILIZING SOLUTION; IT MAY BE THAT THE DEFECT REPORTED HAS BEEN CAUSED BY A FAULT DURING THE MANUFACTURING PROCESS. THIS DEFECT HAS ALREADY BEEN IDENTIFIED AND THE NECESSARY CORRECTIVE MEASURES TAKEN AND TO DATE NO MORE REPORTS HAVE BEEN SUBMITTED FOR THIS REASON, WHICH SHOWS THAT THE ACTIONS HAVE BEEN EFFECTIVE. CONCLUSIONS: THE CLAIM IS DETERMINED AS "CONFIRMED", SINCE THE ATTACHED IMAGE SHOWS THAT IT DOES NOT MEET THE SPECIFICATIONS FOR THIS TYPE OF SUTURE. ACTIONS: NO ACTION IS REQUIRED, THE LEAK IS A DEFECT IDENTIFIED AND ADDRESSED IN CAPA (B)(4)

Description of Event or Problem · 0

COUNTRY OF COMPLAINT: CHILE. THE LIQUID IS FILTERED TO PRESERVE THE SUTURE AND THE CONTAINER IS STAINED. SO THERE ARE SOME SUTURES WITH VERY LITTLE AMOUNT OF LIQUID INSIDE, WHICH IS DOUBTFUL OF THE CORRECT PRESERVATION OF THE SUTURE AND ITS STERILITY; ALSO THE LITTLE AMOUNT OF LIQUID MAKES THE CATGUT VERY RIGID AND MAKES ITS MANEUVERABILITY DIFFICULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
7009 CATGUT PLAIN 3/0 (3) 75CM HR26 SUTURES GAL B.BRAUN SURGICAL SA B0530425 515092

Patients

Seq Age Sex Outcome Treatment
1 Other