FDA Adverse Event Malfunction Summary report: N

BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE

MDR report key: 7165364 · Received January 4, 2018

Report

Report Number
1213809-2017-00423
Event Type
Malfunction
Date Received
January 4, 2018
Date of Event
December 13, 2017
Report Date
March 9, 2018
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
MEG
UDI-DI
30382903052708
PMA / PMN Number
K011103
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SAMPLE EVALUATION: ONE LOOSE 3ML INTEGRA ASSEMBLED SYRINGE WITH NEEDLE WAS RECEIVED BY BD (B)(4) AND REPORTED TO BE FROM BATCH #4202864. THE SAMPLE WAS VISUALLY EVALUATED. THE SYRINGE WAS NOT ACTIVATED AND THE STOPPER WAS IDENTIFIED TO BE FROM MOLD C-249 CAVITY 26. THE STOPPER APPEARS TO HAVE BEEN CUT IN FOUR LOCATIONS BY THE CUTTER. THESE CUTS IN THE STOPPER APPEAR TO BE ALL THE WAY THROUGH THE STOPPER WHICH WOULD RESULT IN LEAKAGE. DHR REVIEW FOR BATCH 4202864: MANUFACTURING DATE: 8/03/2014 TO 08/06/2014. BATCH QUANTITY WAS 360,000. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 4202864 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. INVESTIGATION CONCLUSION: BASED ON THE SAMPLE EVALUATION: CONFIRMED: BD (B)(4) WAS ABLE TO CONFIRM THE CUSTOMER'S INDICATED FAILURE. ROOT CAUSE UNDETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A BD INTEGRA¿ SYRINGE WITH DETACHABLE NEEDLE LEAKED DURING USE. THERE WAS NO REPORT OF EXPOSURE TO MUCOUS MEMBRANES, INJURY OR MEDICAL INTERVENTIONS.