BD LUER-LOK¿ TIP SYRINGE
Report
- Report Number
- 1213809-2017-00409
- Event Type
- Malfunction
- Date Received
- January 4, 2018
- Date of Event
- December 11, 2017
- Report Date
- March 9, 2018
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- PMA / PMN Number
- K980987
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. DHR REVIEW FOR BATCH 7212544 (P/N 302995 ): MANUFACTURING DATE: TO 08/21/17. BATCH QUANTITY WAS 504,000. ALL VISUAL INSPECTIONS WERE PERFORMED AS PER REQUIREMENT WITH NO QUALITY NOTIFICATIONS RELATED TO THE COMPLAINT DEFECT. BATCH 7212544 WAS INSPECTED AND ACCEPTED BASED ON MEETING OUR INSPECTION CONTROL PLAN AND SUBSEQUENTLY APPROVED FOR SHIPMENT. PLEASE NOTE THAT SILICONE IS AN INERT, NON-TOXIC MEDICAL SUBSTANCE USED AS A LUBRICANT FOR DISPOSABLE HYPODERMIC PRODUCTS. IT IS AN INTEGRAL PART OF THE SYRINGE, ENABLING IT TO PERFORM AS REQUIRED IN VARIOUS CLINICAL APPLICATIONS AND DOES NOT PRESENT A SAFETY OR EFFICACY ISSUE NOR DOES IT IMPACT PRODUCT FUNCTION. THE SILICONE APPLICATION PROCESS IS DESIGNED TO PROVIDE AN EVEN DISTRIBUTION OF SILICONE ON THE INTERIOR OF THE SYRINGE BARREL. SILICONE HAS BEEN IN USE IN THIS APPLICATION FOR OVER 20 YEARS. NO REPORTS ARE KNOWN OF ADVERSE CLINICAL EFFECTS ASSOCIATED WITH THESE PRODUCTS AND UNINTENTIONAL DELIVERY OF SILICONE FLUID LUBRICANT. INVESTIGATION CONCLUSION: UNABLE TO DETERMINE ROOT CAUSE, NO DEFECTS WERE CONFIRMED.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. FDA NOTIFIED?: THE INITIAL REPORTER ALSO NOTIFIED THE FDA VIA MEDWATCH # MW5073363. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BEFORE USE, FOREIGN MATTER WAS FOUND IN A BD LUER-LOK¿ TIP SYRINGE. THE OILY SUBSTANCE WAS FOUND WITHIN THE SYRINGE SHAFT. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10330 | BD LUER-LOK¿ TIP SYRINGE | SYRINGE WITH NEEDLE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 7212544 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |