FDA Adverse Event
Malfunction
Summary report: N
GLUCOMMANDER
MDR report key: 7163932
·
Received January 4, 2018
Report
- Report Number
- 3005853093-2018-00001
- Event Type
- Malfunction
- Date Received
- January 4, 2018
- Date of Event
- December 8, 2017
- Report Date
- January 4, 2018
- Manufacturer
- GLYTEC, LLC
- Product Code
- NDC
- UDI-DI
- 00860057000305
- PMA / PMN Number
- K113853
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A PATCH WAS CREATED IN VERSION 3.3.4.5, WHICH WAS RELEASED JANUARY 3, 2018. THE EFFECTED CLIENTS WILL BE UPGRADED AS SOON AS POSSIBLE.
Description of Event or Problem · 1
AFTER A SPLIT TRANSITION, EDITING THE ORDER SET RENDERS A FULL DOSE INSTEAD OF SECOND HALF OF SPLIT DOSE AT NEXT BASAL TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 9556 | GLUCOMMANDER | PREDICTIVE PULMONARY FUNCTION VALUE CALCULATOR | NDC | GLYTEC, LLC | 3.3.2.0 | 00860057000305 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |