FDA Adverse Event
Malfunction
Summary report: N
X SERIES
MDR report key: 7163370
·
Received January 4, 2018
Report
- Report Number
- 1220908-2017-03391
- Event Type
- Malfunction
- Date Received
- January 4, 2018
- Report Date
- December 15, 2017
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- UDI-DI
- 00847946012663
- PMA / PMN Number
- K112432
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED FLEX CABLE. THE FLEX CABLE WAS RESEATED TO RESOLVE THE MALFUNCTION. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE FAILED SELF-TEST. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 9525 | X SERIES | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | X SERIES | NA | 00847946012663 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |