FDA Adverse Event Injury Summary report: N

MECTACER CERAMIC BALL HEAD

MDR report key: 7163363 · Received January 4, 2018

Report

Report Number
3005180920-2017-00817
Event Type
Injury
Date Received
January 4, 2018
Date of Event
December 8, 2017
Report Date
February 2, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809200
PMA / PMN Number
K112115
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 02 FEBRUARY 2018 R&D PRODUCT MANAGER ANALYZED THE ITEM RECEIVED AND COMMENTED: THE CERAMIC BALL HEAD WAS ANALYZED BUT DID NOT SHOW ANY PARTICULAR SIGN OF FAILURE OF THE DEVICE; SOME SIGNS DUE TO THE REVISION WERE NOTICED ON THE SURFACE AND SOME YELLOW AREAS, PROBABLY SOLIDIFIED ADIPOSE TISSUE OCCURRED DURING REVISION SURGERY, WAS PRESENT IN SOME SMALL AREAS. THE ROOT CAUSE OF THE EVENT CANNOT BE DETERMINED.

Description of Event or Problem · 0

THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE PATHOGEN IS NOT AVAILABLE. THE SURGEON SWAPPED THE HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON (B)(6) 2018: LOT 172270: 102 ITEMS MANUFACTURED AND RELEASED ON (B)(6) 2017. EXPIRATION DATE: (B)(6) 2022 NO ANOMALIES FOUND RELATED TO THE ISSUE. TO DATE, 72 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN DUE TO SIGNS OF INFECTION. THE PATHOGEN IS NOT AVAILABLE. THE SURGEON SWAPPED THE HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9329 MECTACER CERAMIC BALL HEAD CERAMIC FEMORAL HEAD LZO MEDACTA INTERNATIONAL SA 172270 07630030809200

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention