FDA Adverse Event
Other
Summary report: N
BD LOGIC BLOOD GLUCOSE MONITOR
MDR report key: 716076
·
Received May 18, 2006
Report
- Report Number
- 2243072-2006-00198
- Event Type
- Other
- Date Received
- May 18, 2006
- Date of Event
- April 24, 2006
- Report Date
- May 18, 2006
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT HER METER IS READING HIGH. ANALYSIS RUN ON CONSESUS ERROR GRID (CEG) USING MEAN READINGS (99) AS REFERENCE POINT VS. BD READINGS (170, 300, 210) YIELDED SOME DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD LOGIC BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 4064264 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |