FDA Adverse Event Injury Summary report: N

G7 10 DEG E1 LINER 32MM F

MDR report key: 7159548 · Received January 3, 2018

Report

Report Number
0001825034-2017-11486
Event Type
Injury
Date Received
January 3, 2018
Date of Event
December 14, 2017
Report Date
January 23, 2018
Manufacturer
ZIMMER BIOMET, INC.
Product Code
PBI
PMA / PMN Number
PK121874
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: 650-0834, DELTA CER FEM HD 032/0MM 12/14, 2017091003; 010000665, G7 PPS LTD ACET SHELL 56F, 6143040. REPORT SOURCE, FOREIGN ¿ EVENTS OCCURRED IN (B)(6). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. PRODUCT LOCATION UNKNOWN.

Additional Manufacturer Narrative · 1

(B)(4). THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY ADDITIONAL INFORMATION WHICH WAS UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS CONFIRMED THROUGH REVIEW OF THIRD PARTY X-RAY REPORT. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS NO TRENDS WERE IDENTIFIED. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS NON-COMPLIANT AND EXPERIENCED MULTIPLE DISLOCATION EVENTS. A REVISION PROCEDURE HAS BEEN SCHEDULED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO MULTIPLE DISLOCATION EVENTS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD A LEFT TOTAL HIP REVISION SURGERY APPROXIMATELY 15 DAYS POST PRIMARY SURGERY, DUE TO TWICE OF DISLOCATIONS. DURING THE REVISION, THE LINER AND HEAD WERE REPLACED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4199 G7 10 DEG E1 LINER 32MM F PROSTHESIS, HIP PBI ZIMMER BIOMET, INC. N/A 6014247

Patients

Seq Age Sex Outcome Treatment
1 56 YR Required Intervention