FREESTYLE LIBRE
Report
- Report Number
- 2954323-2017-09209
- Event Type
- Injury
- Date Received
- December 29, 2017
- Date of Event
- December 6, 2017
- Report Date
- January 15, 2019
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- PZE
- PMA / PMN Number
- P160030
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- 003
Narratives
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. THE DATE OF EVENT IS UNKNOWN. THE MFG DATE IS UNKNOWN. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
NO PRODUCT HAS BEEN RETURNED. EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND DETERMINED THAT THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. THE REPORTED COMPLAINT IS RELATED TO SKIN IRRITATION OR AN ALLERGIC REACTION TO THE PATCH ADHESIVE OF THE FREESTYLE LIBRE SENSOR. DOSE AUDIT AND ENVIRONMENTAL MONITORING REPORTS WERE REVIEWED FOR ISSUES RELATING TO STERILITY OF THE PRODUCT. THE SENSOR COMPONENT HAS NO IMPACT ON PRODUCT STERILITY AND THEREFORE, SENSOR COMPONENTS DHRS WERE NOT REVIEWED. SENSOR KIT DHRS HAVE BEEN REVIEWED TO ASSESS THE MANUFACTURING PROCESS, WHICH INCLUDES THE APPLICATION OF THE ADHESIVE TO THE PUCK. A DHR (DEVICE HISTORY REVIEW) FOR THE FREESTYLE LIBRE SENSOR KIT WAS REVIEWED AND THE DHR SHOWED THE FREESTYLE LIBRE SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. DOSE AUDIT REPORTS WERE REVIEWED AND DEMONSTRATES THE CONTINUED EFFECTIVENESS OF THE ESTABLISHED STERILIZATION PROCESS FOR LIBRE SENSOR KITS. ENVIRONMENTAL MONITORING REPORTS WERE REVIEWED, INCLUDING BIOBURDEN AND ENDOTOXIN TESTING, AND DEMONSTRATED THAT ALL MONITORING PROCESSES CONTINUE TO MEET ADC MINIMUM REQUIREMENTS FOR PRODUCT QUALITY. THE DOSE AUDIT REPORT COVERS THE REQUIRED TIME PERIOD. ALL REVIEW ACTIVITIES CONDUCTED ABOVE, INCLUDING BUT NOT LIMITED TO THE FINAL RELEASE TESTING SPECIFICALLY ASSOCIATED WITH THE MANUFACTURE OF THIS PRODUCT, ARE SUFFICIENT INFORMATION IN ORDER TO SHOW IF THE PRODUCT HAS MET SPECIFICATIONS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, THE CASE WILL BE RE-OPENED AND A PHYSICAL INVESTIGATION WILL BE PERFORMED.
AN ALLERGIC REACTION WAS REPORTED WITH USE OF AN ADC FREESTYLE LIBRE SENSOR. CUSTOMER HAD CONTACT WITH A DOCTOR AND WAS PRESCRIBED AN UNSPECIFIED OINTMENT. NO ADDITIONAL DETAILS WERE PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
AN ALLERGIC REACTION WAS REPORTED WITH USE OF AN ADC FREESTYLE LIBRE SENSOR. CUSTOMER HAD CONTACT WITH A DOCTOR AND WAS PRESCRIBED AN UNSPECIFIED OINTMENT. NO ADDITIONAL DETAILS WERE PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 933643 | FREESTYLE LIBRE | FLASH GLUCOSE MONITORING SYSTEM | PZE | ABBOTT DIABETES CARE INC. | 71702-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |