CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Report
- Report Number
- 9611109-2017-01056
- Event Type
- Malfunction
- Date Received
- December 29, 2017
- Date of Event
- December 5, 2017
- Report Date
- May 18, 2018
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWA
- PMA / PMN Number
- K032213
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
LIVANOVA (B)(4) MANUFACTURES THE CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP. THE INCIDENT OCCURRED IN (B)(6) A LOAN DRIVE UNIT HAS BEEN PROVIDED TO THE CUSTOMER. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE.IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. DEVICE NOT RETURNED.
A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE. THE SERVICE REPRESENTATIVE WAS ABLE TO CONFIRM THE REPORTED ISSUE AND TRACED THE FAILURE TO A DEFECTIVE ATTACHMENT ARM. THE TECHNICIAN REPLACED THE PART AND ALSO RETIGHTENED THE CARRIAGE FIXING POST TO RESOLVE THE REPORTED ISSUE. SUBSEQUENT FUNCTIONAL VERIFICATION TESTING WAS COMPLETED WITHOUT FURTHER ISSUES AND THE UNIT WAS RETURNED TO SERVICE.
LIVANOVA (B)(4) RECEIVED A REPORT THAT THE CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP STOPPED CIRCULATING AND AN DISPLAYED AN ERROR MESSAGE DURING A PROCEDURE. THE USER SWITCHED OUT THE DRIVE UNIT TO CONTINUE THE CASE. THE REPLACED DRIVE UNIT WAS LATER TESTED ON A DIFFERENT CONSOLE AND THE SAME ERROR MESSAGE WAS DISPLAYED. THERE WAS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 934609 | CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP | CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS | DWA | LIVANOVA DEUTSCHLAND | 60-01-00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |