FDA Adverse Event Injury Summary report: N

TI LUMBAR EXTENSION SIZE 13/220MM RADIUS

MDR report key: 7154235 · Received December 29, 2017

Report

Report Number
8030965-2017-51008
Event Type
Injury
Date Received
December 29, 2017
Date of Event
October 25, 2017
Report Date
December 1, 2017
Manufacturer
OBERDORF : SYNTHES PRODUKTIONS GMBH
Product Code
MDI
UDI-DI
07611819747623
PMA / PMN Number
K142587
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED ON PART NUMBER: 497.254, LOT NUMBER: H193627: MANUFACTURING SITE: (B)94), SUPPLIER DHR: N/A, RELEASE TO WAREHOUSE DATE: 04-OCT-2016: NO NON- CONFORMANCE REPORTS (NCRS) WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. CORRECTED DATA: PHYSICAL MANUFACTURER. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PATIENT WEIGHT IS UNKNOWN. EVENT DATE REPORTED AS (B)(6) 2017. IT IS UNKNOWN IF THIS IS THE DATE THE DEVICE BROKEN OR THE DATE IT WAS DISCOVERED BROKEN. UNKNOWN. UNKNOWN. COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PRODUCT INVESTIGATION WAS PERFORMED. THE RETURNED LUMBAR EXTENSION ROD IS A COMPONENT OF THE VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB (VEPTR) SYSTEM INSTRUMENT AND IMPLANT SET (146.995). THE VEPTR MECHANICALLY STABILIZES AND DISTRACTS THE THORAX, CORRECTING THORACIC DEFORMITIES AND PROVIDING IMPROVEMENT IN VOLUME OF RESPIRATION AND LUNG GROWTH IN INFANTILE AND JUVENILE PATIENTS DIAGNOSED WITH OR AT RISK OF DEVELOPING THORACIC INSUFFICIENCY SYNDROME PER RELEVANT TECHNIQUE GUIDE. THE LUMBAR EXTENSION IS UTILIZED FOR RIB-TO-LUMBAR LAMINA, RIB-TO-LUMBAR OR RIB-TO-ILIUM CONSTRUCTS AND IS CONNECTED TO A RIB SLEEVE BY AN INFERIOR DISTRACTION LOCK. THE RETURNED LUMBAR EXTENSION ROD WAS EXAMINED AND THE COMPLAINT CONDITION WAS ABLE TO BE CONFIRMED AS THE ROD IS BROKEN IN TWO LOCATIONS. AT THE FIRST HOLE AND ALSO ON THE SOLID CYLINDRICAL PORTION OF THE ROD APPROXIMATELY 27MM (CALIPERS CA592) FROM THE TRANSITION OF ROD GEOMETRIES. THE MATERIAL SURFACE AT THE FRACTURE SITES APPEAR HOMOGENEOUS WHEN VIEWED UNDER 5X MAGNIFICATION. THERE IS A PORTION OF THE ROD THAT IS MISSING/WAS NOT RETURNED. THE REMAINDER OF THE DEVICE SHOWS WEAR CONSISTENT WITH IMPLANTATION AND REMOVAL. NO DEFINITIVE ROOT CAUSE WAS ABLE TO BE DETERMINED WITH THE PROVIDED INFORMATION. BECAUSE IT WAS REPORTED THAT THIS IS NOW THE THIRD ROD TO BREAK AT THE SAME POSITION IN THE SAME PATIENT, IT IS MOST LIKELY THAT THE ROD SIZE AND/OR CONSTRUCT IS NOT ADEQUATE FOR THE PATIENT NEEDS OR THAT PATIENT NON-COMPLIANCE HAS POSSIBLY OCCURRED. SIMILARLY, A MATERIAL CHECK WAS NOT PERFORMED AT CQ AS THERE IS NO INDICATION THAT THE MATERIAL WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. WHETHER THIS COMPLAINT CAN BE REPLICATED AT CUSTOMER QUALITY (CQ) IS NOT APPLICABLE FOR THIS COMPLAINT CONDITION AS THE RETURNED DEVICE IS ALREADY BROKEN. NO NEW MALFUNCTIONS WERE IDENTIFIED AS A RESULT OF THE INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORDS WAS UNABLE TO BE PERFORMED SINCE THE LOT NUMBER WAS UNKNOWN. THE PORTION OF THE PLATE WITH THE LOT# WAS A PORTION THAT BROKE OFF AND WAS NOT RETURNED. THE ONLY REMAINING ETCHES ON THE RETURNED PART ARE "13 SYNTHES LOGO AND 497". THEREFORE, WITH THE "13" ETCH BEFORE THE SYNTHES LOGO ETCH, IT CAN BE DETERMINED THAT THE RETURNED ROD IS ACTUALLY PART# 497.254, RATHER THAN THE REPORTED PART# 497.252 BY REVIEW OF TABULATED PRODUCT DRAWING 497_131 CURRENT REVISION H. AS A RESULT, THE PART# IN THIS COMPLAINT FILE WAS UPDATED ON 26-FEB-2018 FROM 497.252 TO 497.254 TO ACCOUNT FOR THE RETURNED DEVICE THAT IS BROKEN. TABULATED PRODUCT DRAWING FOR THE FAMILY OF LONG TITANIUM LUMBAR EXTENSIONS RODS WAS REVIEWED DURING THIS INVESTIGATION. THE DESIGN, MATERIALS AND FINISHING PROCESSES WERE FOUND TO BE APPROPRIATE FOR THE INTENDED USE OF THIS DEVICE. FOR THE RETURNED DEVICE, THE DIAMETER OF THE CYLINDRICAL PORTION OF THE ROD NEAR LOCATION OF BREAKAGE MEASURED AT CQ AND IS WITHIN SPECIFICATION PER TABULATED PRODUCT DRAWING. FOR THE RETURNED DEVICE, THE HEIGHT OF THE "T" SHAPED PORTION OF THE ROD NEAR LOCATION OF BREAKAGE MEASURED AT CQ AND IT IS WITHIN SPECIFICATION PER TABULATED PRODUCT DRAWING WITH REFERENCE TO GENERAL TOLERANCES FOR DESIGN AND MACHINING WORK INSTRUCTION. FOR THE RETURNED DEVICE, THE MAJOR WIDTH OF THE "T" SHAPED PORTION OF THE ROD NEAR LOCATION OF BREAKAGE MEASURED AT CQ AND IS WITHIN SPECIFICATION PER TABULATED PRODUCT DRAWING. FOR THE RETURNED DEVICE, THE MINOR WIDTH OF THE "T" SHAPED PORTION OF THE ROD NEAR LOCATION OF BREAKAGE MEASURED AT CQ AND IS WITHIN SPECIFICATION PER TABULATED PRODUCT DRAWING WITH REFERENCE TO GENERAL TOLERANCES FOR DESIGN AND MACHINING WORK INSTRUCTION. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PLATE BROKE AT THE FIRST HOLE. THIS IS NOW THE THIRD PLATE BREAK, AT THE SAME POSITION IN THE SAME PATIENT. NO INFORMATION ABOUT PATIENT STATUS. SEE ALSO COMPLAINT (B)(4). THIS COMPLAINT INVOLVES 1 PART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
934462 TI LUMBAR EXTENSION SIZE 13/220MM RADIUS PROSTHESIS, RIB REPLACEMENT MDI OBERDORF : SYNTHES PRODUKTIONS GMBH H193627 07611819747623

Patients

Seq Age Sex Outcome Treatment
1 13 YR Required Intervention