VGXP XP MED TIB BRG LT 9X63
Report
- Report Number
- 0001825034-2017-10913
- Event Type
- Injury
- Date Received
- December 28, 2017
- Date of Event
- March 18, 2016
- Report Date
- December 5, 2017
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- MBH
- PMA / PMN Number
- PK132873
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). CONCOMITANT MEDICAL PRODUCTS: CAT#195918; LOT#498450 - VGXP INTLK FEMORAL LT 55. CAT#195751; LOT#833150 - VANG XP INLK PRI TIB TRAY 63 MM. CAT#195807; LOT#759920 - VGXP XP LAT TIB BRG LT 9X63. RELATED MFR REPORTS: 0001825034-2017-10910, 0001825034-2017-10911, 0001825034-2017-10912. PRODUCT WAS NOT RETURNED AND NO PHOTOS OF THE DEVICE WERE PROVIDED; THEREFORE, NO EVALUATIONS COULD BE CONDUCTED. REVIEW OF DEVICE HISTORY RECORDS FOR PN 195877 LN 891500 FOUND THESE UNITS WERE RELEASED TO DISTRIBUTION WITH NO DEVIATIONS OR ANOMALIES. THE REPORTED DEVICES ARE USED FOR TREATMENT. THE FOLLOWING COMPONENTS WERE REVIEWED FOR COMPATIBILITY WITH NO ISSUES NOTED. PN 195918 LN 498450, PN 195751 LN 833150, PN 195807 LN 759920 , PN 195877 LN 891500. A COMPLAINT HISTORY REVIEW WAS PERFORMED FOR PART #195877. THE SEARCH IDENTIFIED NO ADDITIONAL COMPLAINTS FOR THE SAME OR SIMILAR ISSUE OF FRACTURE. THE SEARCH ALSO IDENTIFIED 4 OTHER COMPLAINTS FOR THE LOT (891500) WITH PAIN. REVIEW OF THE OPERATIVE RECORD INDICATES THAT THE SURGEON NOTED TIBIAL ISLAND DIDN'T STAY INTACT THROUGH FULL EXTENSION DURING TRIALING. SURGICAL NOTES WERE NOT PROVIDED; IT IS UNKNOWN WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. WITHOUT THE OPPORTUNITY TO EXAMINE THE COMPLAINT PRODUCT, ROOT CAUSE CANNOT BE DETERMINED NO CORRECTIVE ACTIONS, PREVENTIVE ACTIONS, OR FIELD ACTIONS RESULTED AFTER INVESTIGATION OF THIS EVENT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LEFT KNEE SURGERY ON (B)(6) 2016. THE OPERATIVE RECORD INDICATED TIBIAL ISLAND DID NOT STAY INTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 931633 | VGXP XP MED TIB BRG LT 9X63 | PROSTHESIS - KNEE | MBH | ZIMMER BIOMET, INC. | N/A | 891500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Other |