FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 7150244 · Received December 27, 2017

Report

Report Number
3004753838-2017-110001
Event Type
Malfunction
Date Received
December 27, 2017
Date of Event
November 28, 2017
Report Date
November 28, 2017
Manufacturer
DEXCOM, INC.
Product Code
PQF
UDI-DI
10386270000221
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT THE SMART DEVICE SHOWED A TRANSMITTER FAILED ERROR ON (B)(6) 2017 . NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE TRANSMITTER HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. VOLTAGE TEST FAILED. SHARE LOG REVIEW SHOWED TRANSMITTER FAILURE ALERT IN LOG. THE COMPLAINT WAS CONFIRMED BECAUSE A TRANSMITTER ERROR ALERT IN THE CLOUD DATA THE REPORTED EVENT OF A FAILED TRANSMITTER ERROR WAS CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED. 1. PROBLEM NOT CONFIRMED. INVESTIGATION SUMMARY:1. PERFORM EXTERIOR VISUAL INSPECTION: PASSED 2. VOLTAGE MEASUREMENT: FAILED 0.21 VDC 3. PERFORM PAIRING TEST WITH NORDIC BLUETOOTH DEVICE: PASSED 4. DATA AVAILABLE IN THE CLOUD\SHARE AND PERFORM REVIEW OF THIS DATA: YES - TRANSMITTER ERROR WAS NOT FOUND 5. PERFORM REVIEW OF THE BIN FILE: YES-TRANSMITTER ERROR WAS NOT FOUND FFA LAB PRODUCT AND DATA INVESTIGATION I. FFA LAB COULD NOT FIND\CONFIRM A TRANSMITTER FAILED ERROR WITHIN THE INVESTIGATION WINDOW USING THE DEXCOM SHARE SUPPORT TOOL. II. DURING THE DATA BASE BIN FILE INVESTIGATION IN THE LOG INFORMATION TAB - SHOWS THAT NO "BATT CRT" IN THE LOG INFORMATION COLUMN WAS NEVER FOUND AND THIS CONFIRMS THAT THE TRANSMITTER NEVER SENT THE "TRANSMITTER FAILED ERROR" OR "TRANSMITTER FATAL ERROR" MESSAGE TO THE SMART DEVICE AND\OR RECEIVER. PLEASE SEE THE ATTACHED FILE "FFA LAB PRODUCT AND DATA INVESTIGATION TX UNIT 413AYJ FOR SR-(B)(4)" III. CUSTOMER COMPLAINT CANNOT BE CONFIRMED WITH TRANSMITTER'S HARDWARE TEST RESULTS ROOT CAUSE:THE CUSTOMER ALLEGATION WAS NOT FOUND

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
927812 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR PQF DEXCOM, INC. 9438-06 6002976 10386270000221

Patients

Seq Age Sex Outcome Treatment
1 9 YR