DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2017-110001
- Event Type
- Malfunction
- Date Received
- December 27, 2017
- Date of Event
- November 28, 2017
- Report Date
- November 28, 2017
- Manufacturer
- DEXCOM, INC.
- Product Code
- PQF
- UDI-DI
- 10386270000221
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4)
(B)(4).
DEXCOM WAS MADE AWARE ON (B)(4) 2017 THAT THE SMART DEVICE SHOWED A TRANSMITTER FAILED ERROR ON (B)(6) 2017 . NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE TRANSMITTER HAS BEEN RECEIVED FOR EVALUATION. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION HAS BEEN COMPLETED.
THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. VOLTAGE TEST FAILED. SHARE LOG REVIEW SHOWED TRANSMITTER FAILURE ALERT IN LOG. THE COMPLAINT WAS CONFIRMED BECAUSE A TRANSMITTER ERROR ALERT IN THE CLOUD DATA THE REPORTED EVENT OF A FAILED TRANSMITTER ERROR WAS CONFIRMED. THE ROOT CAUSE COULD NOT BE DETERMINED. 1. PROBLEM NOT CONFIRMED. INVESTIGATION SUMMARY:1. PERFORM EXTERIOR VISUAL INSPECTION: PASSED 2. VOLTAGE MEASUREMENT: FAILED 0.21 VDC 3. PERFORM PAIRING TEST WITH NORDIC BLUETOOTH DEVICE: PASSED 4. DATA AVAILABLE IN THE CLOUD\SHARE AND PERFORM REVIEW OF THIS DATA: YES - TRANSMITTER ERROR WAS NOT FOUND 5. PERFORM REVIEW OF THE BIN FILE: YES-TRANSMITTER ERROR WAS NOT FOUND FFA LAB PRODUCT AND DATA INVESTIGATION I. FFA LAB COULD NOT FIND\CONFIRM A TRANSMITTER FAILED ERROR WITHIN THE INVESTIGATION WINDOW USING THE DEXCOM SHARE SUPPORT TOOL. II. DURING THE DATA BASE BIN FILE INVESTIGATION IN THE LOG INFORMATION TAB - SHOWS THAT NO "BATT CRT" IN THE LOG INFORMATION COLUMN WAS NEVER FOUND AND THIS CONFIRMS THAT THE TRANSMITTER NEVER SENT THE "TRANSMITTER FAILED ERROR" OR "TRANSMITTER FATAL ERROR" MESSAGE TO THE SMART DEVICE AND\OR RECEIVER. PLEASE SEE THE ATTACHED FILE "FFA LAB PRODUCT AND DATA INVESTIGATION TX UNIT 413AYJ FOR SR-(B)(4)" III. CUSTOMER COMPLAINT CANNOT BE CONFIRMED WITH TRANSMITTER'S HARDWARE TEST RESULTS ROOT CAUSE:THE CUSTOMER ALLEGATION WAS NOT FOUND
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 927812 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | PQF | DEXCOM, INC. | 9438-06 | 6002976 | 10386270000221 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |