FDA Adverse Event Injury Summary report: N

FREESTYLE HANDS-FREE BREASTPUMP

MDR report key: 7148995 · Received December 27, 2017

Report

Report Number
1419937-2017-00363
Event Type
Injury
Date Received
December 27, 2017
Date of Event
December 8, 2017
Report Date
December 27, 2017
Manufacturer
MEDELA AG
Product Code
HGX
PMA / PMN Number
K150499
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

REPLACEMENT SPARE PARTS WERE SENT TO THE CUSTOMER. IN FOLLOW UP WITH A COMPLAINT HANDLER ON 12/27/2017, THE CUSTOMER CONFIRMED THAT SHE WAS DIAGNOSED AND TREATED FOR A YEAST INFECTION. SHE INDICATED THAT THE REPLACEMENT PARTS WERE WORKING WITHOUT ISSUE AND THAT THE YEAST INFECTION HAS SINCE RESOLVED. IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S THRUSH. REPORTED ISSUES OF THRUSH WERE INVESTIGATED UNDER (B)(4) AND THE ROOT CAUSES WERE IDENTIFIED AS: LACK OF EDUCATION/TRAINING TO MOTHERS ON BREAST FEEDING AND BREAST MILK PUMPING; INSUFFICIENT GUIDANCE ON BREAST SHIELD SIZING TO PREVENT NIPPLE TRAUMA; INSUFFICIENT PERSONAL HYGIENE PRACTICES PRIOR TO BREAST MILK PUMPING OR BREAST FEEDING; AND NO ACCESS TO OR NOT FOLLOWING THE MANUFACTURES' INSTRUCTIONS FOR THE CLEANING AND SANITATION OF THE BREAST MILK PUMPING COMPONENTS. NO CORRECTIVE ACTIONS RESULTED AS IT WAS DETERMINED THAT DETAILED INSTRUCTIONS ARE AVAILABLE AND ADEQUATE IN BOTH PRINTED AND ON-LINE FORMATS FOR CLEANING BREAST MILK PUMPING EQUIPMENT, PERSONAL HYGIENE, PROPER BREAST SHIELD SIZING TO PREVENT NIPPLE SORENESS/TRAUMA, WHICH CAN LEAD TO SKIN BREACHES, ALLOWING FOR THE POTENTIAL INTRODUCTION OF YEAST.

Description of Event or Problem · 1

ON (B)(6) 2017, THE CUSTOMER ALLEGED TO MEDELA LLC THAT THE BACK COVER ON THE SPARE PARTS FOR HER FREESTYLE BREAST PUMP ARE CRACKED. SHE INDICATED THAT BOTH SHE AND HER BABY HAVE THRUSH, FOR WHICH SHE RECEIVED PRESCRIPTION MEDICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
927414 FREESTYLE HANDS-FREE BREASTPUMP PUMP, BREAST, POWERED HGX MEDELA AG 67060

Patients

Seq Age Sex Outcome Treatment
1 Other