UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2017-00022
- Event Type
- Malfunction
- Date Received
- December 22, 2017
- Date of Event
- November 30, 2017
- Report Date
- November 30, 2017
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- UDI-DI
- 15099590381646
- PMA / PMN Number
- K140911
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ON (B)(4) 2017 THE FIELD SERVICE ENGINEER (FSE) REPLACED THE SAMPLE ASPIRATION MODULE (SAM) MODULE RESOURCE CONTROLLER (MRC) TO RESOLVE THE BLOOD SAMPLING VALVE (BSV) MECHANICAL DRIVE COMMUNICATION ISSUE; PER PHONE COMMUNICATION WITH THE FSE ON (B)(4) 2017, THIS IS WHEN THE SPARK AND FLAME WAS OBSERVED COMING FROM THE SAM MRC BOARD. THE FSE REPLACED THE AFFECTED SAM MRC BOARD RESOLVING THE ISSUE. PER BECKMAN COULTER PRODUCT COMPLIANCE ENGINEER, THERE IS NO ELECTRICAL HAZARD PRESENT ON AN MRC BOARD. THE BOARDS OPERATES AT 24 VDC, WHICH IS CONSIDERED A NON-HAZARDOUS VOLTAGE PER SAFETY STANDARD IEC 61010-1 AND IS CONSIDERED SELF-EXTINGUISHABLE. WITH REGARDS TO THE SPARK AND SMALL FLAME SEEN ON THE INSTRUMENT, THIS EVENT IS NOT CLASSIFIED AS A FIRE EVENT, AT BEST THIS EVENT IS CONSIDERED A MOMENTARY EVENT CAUSED EITHER BY A LOW VOLTAGE COMPONENT FAILURE, OR BY THE INADVERTENT SHORTING OF THE BOARD DURING TROUBLESHOOTING. IN AN ABUNDANCE OF CAUTION, BEC HAS FILED THE REPORT FOR THIS EVENT BASED ON THE REPORTED VISIBLE FLAME. BEC INTERNAL IDENTIFIER - (B)(6).
THE FIELD SERVICE ENGINEER (FSE) REPORTED OBSERVING A CLEAR SPARK THEN A SMALL FLAME WITH SMOKE AFTER THE INSTALLING THE SAMPLE ASPIRATION MODULE (SAM) MODULE RESOURCE CONTROLLER (MRC) BOARD WHILE TROUBLESHOOTING BLOOD SAMPLING VALVE (BSV) MOVE ERRORS ON THE DXH 800 INSTRUMENT. PER THE FSE THE MAGNITUDE OF THE FLAME WAS VERY SMALL AND CONTAINED WITHIN THE INSTRUMENT. A FIRE EXTINGUISHER WAS NOT NEEDED AND THE INSTRUMENT WAS IMMEDIATELY POWERED OFF. THE LAB WAS NOT EVACUATED. THERE WAS NO REPORTED INJURY AND NO MEDICAL ATTENTION WAS REQUIRED BY THE OPERATOR. ERRONEOUS RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 921636 | UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | N/A | 15099590381646 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |