FDA Adverse Event Injury Summary report: N

GMK SPHERE FEMORAL COMPONENT

MDR report key: 7142937 · Received December 22, 2017

Report

Report Number
3005180920-2017-00766
Event Type
Injury
Date Received
December 22, 2017
Date of Event
November 22, 2017
Report Date
March 23, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030825897
PMA / PMN Number
K121416
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INFORMATION RECEIVED BY THE SALES AGENT ON 22 FEBRUARY 2018 BY THE SALES REPRESENTATIVE REVISION WAS PERFORMED ON THE (B)(6) 2017 AND WAS COMPLETED SUCCESSFULLY BY CHANGING THE FEMORAL COMPONENT. REVISION WAS REQUIRED DUE TO MALALIGNMENT OF THE FEMORAL COMPONENT. THIS WAS DUE TO EXCESS CEMENT MANTLE ON THE MEDIAL SIDE OF THE IMPLANT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR ON 27 FEBRUARY 2018 EARLY PARTIAL REVISION (FEMORAL COMPONENT ONLY, 9 MONTHS) OF CEMENTED TKA. ACCORDING TO REPORT, THE CAUSE IS MISALIGNMENT DUE TO EXCESSIVELY THICK CEMENT MANTLE ON THE MEDIAL SIDE. THIS PROBLEM IS NOT RELATED TO A COMPONENT MALFUNCTION OR DEFECT.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 18 DECEMBER 2017. LOT 152871: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 23 JUNE 2015. EXPIRATION DATE: 2020-05-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE,(B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK SPHERE TIBIAL INSERT FIXED FLEX #6/12 MM L REF. 02.12.0612FL, LOT. 144617 (K121416): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06 OCT 2014. EXPIRATION DATE: 2019-08-31 NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK TIBIAL TRAY FIXED CEMENTED # 6 L REF. 02.07.1206L, LOT. 142707 (K090988): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 29 JULY 2014. EXPIRATION DATE: 2019-06-30 NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. GMK PRIMARY EXTENSION STEM Ø11MM / L 30 MM REF 02.07.F11030 LOT. 153526 (K133630): (B)(4) ITEMS MANUFACTURED AND RELEASED ON 25 JUNE 2015. EXPIRATION DATE: 2018-02-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

PATIENT HAS REPORTED PAIN AND DISCOMFORT. REVISION WILL OCCUR ON THE (B)(6) 2017. MORE DETAILS TO FOLLOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
922094 GMK SPHERE FEMORAL COMPONENT CEMENTED FEMORAL COMPONENT JWH MEDACTA INTERNATIONAL SA 152871 07630030825897

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention