FDA Adverse Event Malfunction Summary report: N

MICROBORE EXTENSION SET

MDR report key: 7142792 · Received December 22, 2017

Report

Report Number
9616066-2017-01596
Event Type
Malfunction
Date Received
December 22, 2017
Report Date
November 29, 2017
Manufacturer
CAREFUSION
Product Code
FPA
PMA / PMN Number
K790108
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER¿S REPORT OF A CRACKED FEMALE LUER WAS CONFIRMED. VISUAL INSPECTION OBSERVED THAT THE FEMALE LUER HAD A VERTICAL HAIR LINE CRACK MEASURING 0.589 INCHES LONG. THE FEMALE LUER WAS INSPECTED UNDER MAGNIFICATION; NO STRESS MARKS WERE OBSERVED. FUNCTIONAL TESTING WAS PERFORMED; A LEAK WAS OBSERVED AS FLUID FLOWED THROUGH THE CRACK ON THE FEMALE LUER. THE ROOT CAUSE OF THE CRACKED FEMALE LUER WAS NOT DETERMINED.

Additional Manufacturer Narrative · 1

MW 5073184. THE AFFECTED PRODUCT HAS BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH INVESTIGATION RESULTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A VECURONIUM INFUSION, A CRACK WAS OBSERVED AT THE NEEDLELESS CONNECTOR WHERE THE SYRINGE AND FEMALE END OF THE EXTENSION TUBING CONNECTED. NO PATIENT HARM WAS REPORTED AND THE LINE WAS CHANGED IMMEDIATELY. RECEIVED A COPY OF THE CUSTOMER'S SUS VOLUNTARY EVENT REPORT FROM THE FDA WHICH STATES, "FIRST INSTANCE, MORPHINE DRIP WAS FOUND BY BEDSIDE NURSE LEAKING. TUBING WAS CRACKED AT THE MAXZERO CONNECTED TO THE SYRINGE. LINE WAS CHANGED PER PROTOCOL BY LT IMMEDIATELY. SECOND INSTANCE ALARIS (CAREFUSION) MICROBORE EXTENSION SET TUBING WITH 0.2 MICRON LOW PROTEIN BINDING FILTER (PRODUCT REF# (B)(4) WITH CRACK AT FEMALE END. THIS WAS CONNECTED TO A NEEDLELESS CONNECTOR (MAXZERO). NOT ABLE TO ISOLATE A LOT NUMBER FOR THIS VAROOM. THIS IS NOW THE THIRD PRODUCT FOUND CRACKED IN THE SAME PLACE. SEE VAROOM #S (B)(4) 2017. BOTH ON (B)(6) 2017. ALL 3 PRODUCTS WILL BE GIVEN TO RISK MGMT FOR F/U WITH COMPANY. PLEASE SEND REPORT FROM COMPANY WHEN THEY ARE FINISHED WITH THEIR INVESTIGATION. NO HARM TO DONE TO PT. MILRINONE WAS INFUSING VIA THIS TUBING FOR THIS PT. TUBING WAS IN-LINE APPROX. 24HRS BEFORE ONTO MEDFUSION PUMP AND FLOOR WAS NOTED. THIRD INSTANCE, BEDSIDE NOTICED VECURONIUM DRIP TUBING CRACKED AT THE MAXZERO CONNECTING TO THE SYRINGE. LINE CHANGED IMMEDIATELY BY LT." ALTHOUGH THE CUSTOMER SELECTED B2, "OTHER SERIOUS" ON THE MEDWATCH, THE CLINICIAN LATER CLARIFIED THAT NO MEDICAL INTERVENTION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
922767 MICROBORE EXTENSION SET SET, ADMINISTRATION, INTRAVASCULAR FPA CAREFUSION 10010570

Patients

Seq Age Sex Outcome Treatment
1