FDA Adverse Event Injury Summary report: N

LIFELINE PERSONAL RESPONSE SYSTEM

MDR report key: 7141825 · Received December 22, 2017

Report

Report Number
1220762-2017-00002
Event Type
Injury
Date Received
December 22, 2017
Report Date
December 20, 2017
Manufacturer
LIFELINE, INC.
Product Code
ILQ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS REPORTED TO FDA PER 1220762-11-15-2017-001-R, THE INVOLVED BUTTON IS PART OF A FIELD ACTION ON THE PART OF THE MANUFACTURER. THE INVOLVED BUTTON DID NOT HAVE THE APPROPRIATE FALL PARAMETERS. IT IS UNCLEAR FROM THE INFORMATION PROVIDED TO THE MANUFACTURER IF THE BUTTON MAY HAVE CONTRIBUTED TO ANY INJURY THE SUBSCRIBER SUSTAINED. THE INVOLVED BUTTON WAS STILL ABLE TO BE DEPRESSED TO SUMMON HELP. PER AUTOALERT HELP BUTTON, INSTRUCTIONS FOR USE (P/N 0940718 REV. 7) P. 10 THE AUTOALERT HELP BUTTON CAN DETECT MOST FALLS, PROVIDING AN ADDED LAYER OF PROTECTION. IF YOU THINK YOU NEED ASSISTANCE, PUSH YOUR HELP BUTTON IMMEDIATELY TO INITIATE THE HELP CALL. PUSHING THE HELP BUTTON GENERATES THE HELP CALL IMMEDIATELY. P. 13 RECOMMENDED USAGE PUSH YOUR AUTOALERT HELP BUTTON ANY TIME YOU NEED HELP. IF YOU FALL AND ARE ABLE TO, YOU SHOULD STILL PUSH THE AUTOALERT BUTTON TO SEND A HELP CALL RIGHT AWAY. PER AUTOALERT HELP BUTTON, INSTRUCTIONS FOR USE, P/N 0940718 REV. 7 P. 5 STATES: "IN CERTAIN SITUATIONS, THE AUTOALERT HELP BUTTON MAY NOT DETECT A FALL. SOME MOVEMENTS MAY NOT REGISTER AS A FALL AND WOULD NOT BE DETECTED. EXAMPLES INCLUDE, BUT ARE NOT LIMITED TO: A GRADUAL SLIDE SUCH AS FROM A SEATED POSITION; LOWERING ONESELF SLOWLY TO THE GROUND (TO BRACE THE IMPACT OF A FALL) , A FALL FROM A HEIGHT OF LESS THAN 20 INCHES (0.5 METERS)."

Description of Event or Problem · 1

THE SUBSCRIBER WAS UNABLE TO PROVIDE A DATE OF THE FALL. SHE STATED SHE HAD FALLEN FORWARD AND WAS UNABLE TO MOVE. SHE STATED SHE WAS ON THE FLOOR FOR 6 HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
920937 LIFELINE PERSONAL RESPONSE SYSTEM SYSTEM, COMMUNICATOR, POWERED ILQ LIFELINE, INC. FD100

Patients

Seq Age Sex Outcome Treatment
1 87 YR Hospitalization