LIFELINE PERSONAL RESPONSE SYSTEM
Report
- Report Number
- 1220762-2017-00001
- Event Type
- Injury
- Date Received
- December 22, 2017
- Date of Event
- October 2, 2017
- Report Date
- December 20, 2017
- Manufacturer
- LIFELINE SYSTEMS, INC.
- Product Code
- ILQ
- UDI-DI
- 00884838084957
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
AS REPORTED TO FDA IN 1220762-11-15-2017-001-R, THE INVOLVED BUTTON WAS AN AFFECTED DEVICE. THE BUTTON DID NOT HAVE THE APPROPRIATE FALL PARAMETERS. IT IS UNCLEAR FROM THE INFORMATION RECEIVED BY THE MANUFACTURER IF THE BUTTON MAY HAVE CONTRIBUTED TO ANY INJURY THE SUBSCRIBER SUSTAINED. PER AUTOALERT HELP BUTTON, INSTRUCTIONS FOR USE (P/N 0940718 REV. 7) P. 10, THE AUTOALERT HELP BUTTON CAN DETECT MOST FALLS, PROVIDING AN ADDED LAYER OF PROTECTION. IF YOU THINK YOU NEED ASSISTANCE, PUSH YOUR HELP BUTTON IMMEDIATELY TO INITIATE THE HELP CALL. PUSHING THE HELP BUTTON GENERATES THE HELP CALL IMMEDIATELY. P. 13 RECOMMENDED USAGE: PUSH YOUR AUTOALERT HELP BUTTON ANY TIME YOU NEED HELP. IF YOU FALL AND ARE ABLE TO, YOU SHOULD STILL PUSH THE AUTOALERT BUTTON TO SEND A HELP CALL RIGHT AWAY. PER AUTOALERT HELP BUTTON, INSTRUCTIONS FOR USE, P/N 0940718 REV. 7 P. 5 STATES: "IN CERTAIN SITUATIONS, THE AUTOALERT HELP BUTTON MAY NOT DETECT A FALL.SOME MOVEMENTS MAY NOT REGISTER AS A FALL AND WOULD NOT BE DETECTED. EXAMPLES INCLUDE, BUT ARE NOT LIMITED TO: A GRADUAL SLIDE SUCH AS FROM A SEATED POSITION; LOWERING ONESELF SLOWLY TO THE GROUND (TO BRACE THE IMPACT OF A FALL) , A FALL FROM A HEIGHT OF LESS THAN 20 INCHES (0.5 METERS)." THE INVOLVED BUTTON WAS STILL ABLE TO BE DEPRESSED TO SUMMON HELP.
THE SUBSCRIBER FELL AND HIT HER HEAD. IT IS UNCLEAR IF SHE HIT HER HEAD AND THEN HAD A STROKE OR IF THE STROKE CAUSED HER TO FALL AND SHE HIT HER HEAD IN THE PROCESS. THIS HAPPENED IN HER BEDROOM THE NIGHT OF THE (B)(6) OR THE MORNING OF THE (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 920713 | LIFELINE PERSONAL RESPONSE SYSTEM | SYSTEM, COMMUNICATION, POWERED | ILQ | LIFELINE SYSTEMS, INC. | FD100 | 00884838084957 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Hospitalization |