FDA Adverse Event Injury Summary report: N

G7 ACETABULAR SCREW

MDR report key: 7139568 · Received December 21, 2017

Report

Report Number
0001825034-2017-11395
Event Type
Injury
Date Received
December 21, 2017
Date of Event
August 15, 2016
Report Date
March 15, 2018
Manufacturer
ZIMMER BIOMET, INC.
Product Code
PBI
PMA / PMN Number
PK121874
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WHICH EXACT SCREW WAS EXPLANTED IS UNKNOWN. POSSIBLE SCREWS INCLUDE: CATALOG#: 010000998, LOT#: 3801672, EXPIRATION DATE: MAY 15, 2026, UDI #:(B)(4), DEVICE MANUFACTURE DATE: MAY 18, 2016. CATALOG#: 010000997, LOT#: 3753689, EXPIRATION DATE: MAR 1, 2026, UDI #: (B)(4), DEVICE MANUFACTURE DATE: MAR 3, 2016. CATALOG#: 010000999, LOT#: 3787070, EXPIRATION DATE: APR 21, 2026, UDI #: (B)(4), DEVICE MANUFACTURE DATE: APR 25, 2016. CATALOG#: 010000998, LOT#: 3797747, EXPIRATION DATE: MAY 11, 2026, UDI #:(B)(4), DEVICE MANUFACTURE DATE: MAY 16, 2016. MEDICAL PRODUCT: G7 ACETABULAR SYSTEM DUAL MOBILITY ACETABULAR LINER, CATALOG#: 110024462, LOT#: 401090. BIOLOX DELTA HIP SYSTEM MODULAR CERAMIC HEAD, CATALOG#: 12-115109, LOT#: 630110. G7 OSSEO TI 3 HOLE SHELL, CATALOG#: 110010243, LOT#: 3795520. ACT ARCTIC E1 HIP BEARING, CATALOG#: EP-200146, LOT#: 678440. TAPERLOC 133 FP TYPE1 PPS, CATALOG#: 51-100080, LOT#: 3755902. REPORTED EVENT WAS CONFIRMED BY REVIEW OF X-RAYS AND OPERATIVE NOTES. DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION(S) IS/ARE REQUIRED. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. A SUMMARY OF THE INVESTIGATION HAS BEEN SENT TO THE COMPLAINANT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2016-04307, 11395, 11396.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS REVISED APPROXIMATELY 3 MONTHS POST-OPERATIVELY DUE TO MIGRATION AND MALALIGNMENT OF THE DUAL MOBILITY LINER. MALALIGNMENT/MIGRATION WERE INITIALLY NOTED ON RADIOGRAPHS TEN DAYS POST-IMPLANTATION. IN INITIAL OPERATIVE NOTES, SURGEON CLEARLY STATES THAT THERE IS A POSSIBILITY OF NEEDING A HEAD/LINER EXCHANGE IN FUTURE IF POSITION OF LINER CHANGES OR SYMPTOMS PRESENT FROM THE HIP. FROM THE REVISION OPERATIVE REPORT: PATIENT HAD SIGNIFICANT TENDINOSIS AND PARTIAL FRAYING. THERE WAS SIGNIFICANT METALLOSIS IN THE HIP JOINT FROM DISPLACEMENT OF METAL LINER FROM THE CUP. DUAL MOBILITY HEAD WAS REMOVED. TRUNNION WAS NOTED TO BE MILDLY SCRATCHED WITH STEM WELL FIXED. ONE OF THE SCREWS WAS PROMINENT IN THE SHELL AND WAS MOST LIKELY WAS THE REASON FOR DISPLACEMENT OF THE LINER, ACCORDING TO SURGEON. THIS SCREW WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
919419 G7 ACETABULAR SCREW PROSTHESIS, HIP PBI ZIMMER BIOMET, INC. N/A SEE H10 NARRATIVE

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R