SAPIEN TRANSCATHETER
Report
- Report Number
- 2015691-2017-04457
- Event Type
- Injury
- Date Received
- December 21, 2017
- Date of Event
- November 27, 2017
- Report Date
- November 27, 2017
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- NPT
- PMA / PMN Number
- P110021
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
PER THE DOCUMENTS, APPROXIMATELY 7 YEARS AND 7 MONTHS AFTER THE IMPLANT OF A 23MM SAPIEN VALVE IN THE AORTIC POSITION, THE ECHO SHOWED DEGENERATION OF THE VALVE DUE TO SEVERE AORTIC STENOSIS WITH THICKENING OF THE PROSTHETIC CUSPS, MODERATE CENTRAL REGURGITATION AND PERIVALVULAR REGURGITATION. A VALVE IN VALVE WAS PERFORMED USING A SAPIEN 3 VALVE. AT THE TIME OF THE REPORT THE PATIENT WAS STABLE. TWO IMAGES WERE PROVIDED OF THE ORIGINAL 23MM SAPIEN VALVE IMPLANT. THE ENTIRE STUDY WAS NOT PROVIDED. POST DEPLOYMENT ASSESSMENT, THE VALVE APPEARED POSITIONED AND EXPANDED WELL. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION AS IT REMAINS IMPLANTED IN THE PATIENT. PER THE INSTRUCTIONS FOR USE (IFU), DEVICE DEGENERATION IS LISTED AS A POTENTIAL RISK ASSOCIATED WITH THE TAVR PROCEDURE. STRUCTURAL VALVE DETERIORATION (SVD) MAY BE MANIFESTED AS STENOSIS WITH THICKENED LEAFLETS. SVD REFERS TO CHANGES INTRINSIC TO THE VALVE, AND CAN INCLUDE FAILURE MODES SUCH AS WEAR, CALCIFICATION, LEAFLET TEAR, STENT CREEP, LEAFLET DISRUPTION, OR LEAFLET RETRACTION. SVD MAY BE MILD AND NOT REQUIRE ANY INTERVENTION OR IT MAY BE MODERATE TO SEVERE. IT CAN CAUSE THE HEART TO WORK HARDER TO EJECT BLOOD FROM THE VENTRICLE. DEPENDING ON THE SEVERITY IT COULD BE AN INDICATION FOR VALVE REPLACEMENT OR MEDICAL INTERVENTION. TISSUE CALCIFICATION IS A VERY COMMON FAILURE MODE. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT YET FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. IN THIS CASE, THERE WAS NO ALLEGATION OR INDICATION A PRODUCT DEFICIENCY CONTRIBUTED TO THIS ADVERSE EVENT. THE CAUSE FOR THE VALVE DEGENERATION COULD NOT BE CONFIRMED, BUT PROGRESSION OF DISEASE MAY BE A CONTRIBUTING FACTOR. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.
PER THE INSTRUCTIONS FOR USE, PARAVALVULAR LEAK (PVL) AND CENTRAL AORTIC INSUFFICIENCY (CAI) ARE POTENTIAL ADVERSE EVENT ASSOCIATED WITH BIOPROSTHETIC HEART VALVES. PARAVALVULAR LEAK REFERS TO BLOOD FLOWING THROUGH A CHANNEL BETWEEN THE STRUCTURE OF THE IMPLANTED VALVE AND THE CARDIAC TISSUE, AS A RESULT OF A LACK OF APPROPRIATE SEALING OF THE VALVE TO THE TARGET SITE. SOME PVL IS NOT UNCOMMON POST DEPLOYMENT. MANY CASES ARE MILD TO MODERATE, AND EITHER RESOLVE OVER TIME OR DO NOT CAUSE SYMPTOMS. OTHERS MAY BE MORE CLINICALLY SIGNIFICANT AND REQUIRE INTERVENTION. THE MECHANISM BEHIND WORSENING OR LATE PVL IS NOT WELL UNDERSTOOD BUT MAY BE RELATED TO CARDIAC REMODELING. CENTRAL REGURGITATION WHICH DEVELOPS PROGRESSIVELY OVER TIME CAN BE DUE TO A NUMBER OF ISSUES INCLUDING PATIENT RELATED FACTORS OR STRUCTURAL VALVE DETERIORATION, INCLUDING CALCIFICATION, NON-CALCIFIC DEGENERATION, LEAFLET THICKENING OR FIBROSIS, OR A COMBINATION OF THESE. REGURGITATION MAY ALSO DEVELOP PROGRESSIVELY IF HOST FIBROTIC TISSUE, OR PANNUS, GROWS ONTO THE BIOPROSTHETIC VALVE. PANNUS, A CAUSE OF NONSTRUCTURAL DYSFUNCTION, MAY INTERFERE WITH FUNCTIONALITY OF THE DEVICE BY RESTRICTING THE LEAFLET MOTION LEADING TO ABNORMAL COAPTATION. CALCIFICATION OF AN IMPLANTED VALVE MAY BE A MANIFESTATION OF STRUCTURAL VALVE DETERIORATION (SVD). THIS TERM REFERS TO CHANGES INTRINSIC TO THE VALVE, AND CAN INCLUDE FAILURE MODES SUCH AS WEAR, CALCIFICATION, LEAFLET TEAR, STENT CREEP, LEAFLET DISRUPTION, OR LEAFLET RETRACTION. THERE ARE CASES OF SVD THAT RESULT IN A COMBINATION OF REGURGITATION AND DEGENERATION. IT MAY BE MILD AND NOT REQUIRE ANY INTERVENTION OR IT MAY BE MODERATE TO SEVERE. IN THESE CASES, IT CAUSES THE HEART TO WORK HARDER TO EJECT BLOOD FROM THE VENTRICLE. DEPENDING ON SEVERITY IT COULD BE AN INDICATION FOR VALVE REPLACEMENT OR MEDICAL INTERVENTION. A VERY COMMON FAILURE MODE IS TISSUE CALCIFICATION. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. IN THIS CASE, AS LIMITED INFORMATION WAS PROVIDED, THERE ARE NO KNOWN RISK FACTORS THAT COULD HAVE CONTRIBUTED TO THE EVENT (END-STAGE RENAL INSUFFICIENCY, DISORDER OF CALCIUM METABOLISM). THERE WAS NO ALLEGATION OR INDICATION A PRODUCT DEFICIENCY CONTRIBUTED TO THIS ADVERSE EVENT. THE ROOT CAUSE OF THIS EVENT COULD NOT BE DETERMINED. THE CAUSE FOR THE PVL AND CAI COULD NOT BE DETERMINED, HOWEVER, MAY BE RELATED TO DISEASE PROGRESSION. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW. NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE REQUIRED.
AS REPORTED BY OUR AFFILIATES IN (B)(6), APPROXIMATELY 7 YEARS POST IMPLANT OF THE 23MM SAPIEN VALVE IN THE AORTIC POSITION, ECHO SHOWED DEGENERATION OF THE VALVE DUE TO MODERATE INTRA-PROSTHETIC FAILURE BY FIBRO-CALCIFICATION AND PARAVALVULAR LEAK. A VALVE IN VALVE WAS PERFORMED USING SAPIEN 3 VALVE. AT THE TIME OF THE REPORT THE PATIENT WAS STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 918050 | SAPIEN TRANSCATHETER | AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED | NPT | EDWARDS LIFESCIENCES | 9000TFX23 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Required Intervention |