FDA Adverse Event Malfunction Summary report: N

ORA SYSTEM 2000

MDR report key: 7138033 · Received December 21, 2017

Report

Report Number
2028159-2017-04570
Event Type
Malfunction
Date Received
December 21, 2017
Report Date
August 27, 2018
Manufacturer
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Product Code
NCF
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM WAS EXAMINED AND THE REPORTED EVENT WAS NOT REPLICATED. THE ABERROMETER WAS SUBSEQUENTLY REPLACED. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. BASED ON QUALITY ASSURANCE ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE SYSTEM WAS FOUND TO MEET SPECIFICATIONS. THEREFORE THE ROOT CAUSE OF THE REPORTED EVENT IS UNABLE TO BE DETERMINED CONCLUSIVELY. (B)(4).

Additional Manufacturer Narrative · 1

THE ABERROMETER WAS RECEIVED AT THE FOR EVALUATION. A VISUAL ASSESSMENT OF THE RETURNED SAMPLE REVEALED DISCOLORATION ON THE ABERROMETER COVER. ADDITIONALLY, A GAP WAS NOTED AT THE ABERROMETER NOSE COVER. THE RETURNED SAMPLE WAS CONNECTED TO A CART AND TESTED, AND ALL MEASUREMENTS TAKEN PER MOCK SURGERY PASSED SPECIFICATION. UNRELATED TO THE REPORTED EVENT, THE WIDE FIELD OF VIEW (WFOV) WAS FOUND TO BE OUT OF SPECIFICATION. WHILE THE FOCUS CAMERA ALSO FAILED MANUFACTURING SPECIFICATION, IT PASSED THE PRODUCT SPECIFICATION. AS NO ISSUES WERE IDENTIFIED WITH THE ABERROMETER PROVIDING ACCURATE LENS POWER RECOMMENDATIONS, THE ROOT CAUSE IS UNABLE TO BE DETERMINED CONCLUSIVELY. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 1

INVESTIGATION, INCLUDING ROOT CAUSE ANALYSIS, IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE SYSTEMS CALCULATIONS WERE OFF. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
916181 ORA SYSTEM 2000 ABERROMETER, OPHTHALMIC NCF ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER NA ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other