FDA Adverse Event Malfunction Summary report: N

SMART HYPER-HYPOTHERMIA

MDR report key: 7137701 · Received December 21, 2017

Report

Report Number
7137701
Event Type
Malfunction
Date Received
December 21, 2017
Date of Event
November 14, 2017
Report Date
December 12, 2017
Manufacturer
CINCINNATI SUB-ZERO PRODUCTS, INC.
Product Code
DWJ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE BLANKET TEMPERATURE WAS SET AT 33.5 DEGREES CELSIUS AND THE ACTUAL BLANKET TEMPERATURE WAS READING 42. INFANT TEMPERATURE REACHED 32.4. THE PHYSICIAN WAS NOTIFIED. RESPIRATORY THERAPY SWITCHED THE MACHINE AND RESTORED THE SETTINGS. THE OLD MACHINE BLANKET TEMPERATURE WAS NOT FLUCTUATING APPROPRIATELY. INFANT TEMPERATURE NOW STABLE AND APPROPRIATE FOR BLANKET SETTING. NO UNTOWARD PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
917746 SMART HYPER-HYPOTHERMIA SYSTEM, THERMAL REGULATING DWJ CINCINNATI SUB-ZERO PRODUCTS, INC.

Patients

Seq Age Sex Outcome Treatment
1 1 YR