FDA Adverse Event Malfunction Summary report: N

BD¿ RELION® INSULIN SYRINGE

MDR report key: 7137173 · Received December 21, 2017

Report

Report Number
1920898-2017-00380
Event Type
Malfunction
Date Received
December 21, 2017
Date of Event
November 1, 2017
Report Date
April 20, 2018
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00681131311786
PMA / PMN Number
K024112
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 7170877 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE THREE NOTIFICATIONS NOTED FOR POOLING. THERE WERE THIRTEEN NOTIFICATIONS NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE SAMPLES / PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: N/A. A SAMPLE IS NOT AVAILABLE FOR EVALUATION. HOWEVER, A NO SAMPLE INVESTIGATION AND DEVICE HISTORY RECORD REVIEW WILL BE COMPLETED. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A BD¿ RELION® INSULIN SYRINGE SEPARATES WHEN TAKING OFF THE CAP. THEY ALSO REPORTED LIQUID FOREIGN MATTER AT THE HUB OF THE SYRINGE. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
917008 BD¿ RELION® INSULIN SYRINGE INSULIN SYRINGE FMF BD MEDICAL - DIABETES CARE 7170877 00681131311786

Patients

Seq Age Sex Outcome Treatment
1 Other