BD SYRINGE WITH NEEDLE
Report
- Report Number
- 3002682307-2017-00128
- Event Type
- Malfunction
- Date Received
- December 20, 2017
- Date of Event
- November 27, 2017
- Report Date
- January 12, 2018
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
WE HAVE BEEN PROVIDED WITH THE AFFECTED SAMPLE. A LEAKAGE THROUGH THE PLUNGER ROD WAS OBSERVED IN THIS PROVIDED SAMPLE. AFTER THAT WE COULD DETERMINE DAMAGE IN THE PLUNGER ROD BY THE EVALUATION OF THE PLUNGER WITH MAGNIFICATION. WE HAVE REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QA NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE N¿2020 (AUGUST 30TH - SEPTEMBER 9TH, 2016). SYRINGES WERE ASSEMBLED IN MACHINE, N¿4255, N¿4254, N¿4237, N¿4220, N¿4210, AND N¿4236, IN LOT #6242444 (AUGUST 29TH - SEPTEMBER 5TH, 2016). RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. WE HAVE ALSO REVIEWED THE BARRELS LOTS #6239273, AND #6230484 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #6239282, AND #6230006 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. CONFIRMATION THE PROVIDED SAMPLE PRESENTED THE REPORTED ISSUE. WE COULD CONFIRM THE REPORTED ISSUE. ROOT CAUSE ANALYSIS WE CONCLUDE THAT THE CAUSE OF THE PROBLEM WAS PRODUCED AS A CONSEQUENCE OF A DAMAGE IN THE PLUNGER LIP. THIS COULD BE PRODUCED DURING THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING PROCESS OR IN THE PLUNGER ASSEMBLY MACHINE.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT DURING USE OF A BD" SYRINGE. A NURSE FOUND SOLUTION LEAKED FROM PLUNGER ROD WHEN WITHDRAWING SOLUTION. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION NEEDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 912236 | BD SYRINGE WITH NEEDLE | SYRINGE AND NEEDLE | FMF | BECTON DICKINSON, S.A. | 1609106 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |