HANDPIECE MICS
Report
- Report Number
- 3005985723-2017-00639
- Event Type
- Malfunction
- Date Received
- December 20, 2017
- Date of Event
- December 12, 2017
- Report Date
- March 6, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K170581
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: THE SCREW IN THE HANDPIECE HANDLE WAS LOOSE/FELL OUT. PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE MANUFACTURED UNDER LOT K09ZR AND (B)(4) INCLUDING 4202549 WERE ACCEPTED INTO FINAL STOCK ON 8/22/2017. A REVIEW OF (B)(4) SHOWS THAT THE ISSUES ARE NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLAINT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO PARTS IN LOT NUMBER K09ZR, P/N 209063 SHOWS NO OTHER COMPLAINT RELATED TO THE FAILURE IN THIS INVESTIGATION. VISUAL INSPECTION: VISUAL INSPECTION REVEALED NO PHYSICAL DAMAGE OF UNIT. THE SCREW WAS OUTSIDE OF THE UNIT. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED VISUAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. MATERIAL ANALYSIS: MATERIAL ANALYSIS WAS NOT COMPLETED BECAUSE THE FAILURE WAS FUNCTIONAL. FUNCTIONAL INSPECTION THE HANDPIECE MOTOR AND ELECTRONICS FUNCTION AS INTENDED. CONCLUSION: THE SCREW HOLDS THE HANDLE IN PLACE. IF THE SCREW BACKS OUT THEN THE HANDLE CAN FALL. NC/CAPA HISTORY REVIEW A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THAT (B)(4) AND CAPA (B)(4) ARE ASSOCIATED WITH THE FAILURE MODE REPORTED IN THIS EVENT. TREND #(B)(4)
SET SCREW FELL OUT OF MICS HANDLE CAUSING THE HANDLE TO FALL OFF. TKA CASE DELAYED FOR 5 MINUTES.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
SET SCREW FELL OUT OF MICS HANDLE CAUSING THE HANDLE TO FALL OFF. TKA CASE DELAYED FOR 5 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 912421 | HANDPIECE MICS | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 4202549 / 42010717 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |