FLEX FOCUS 800 SYSTEM
Report
- Report Number
- 9680269-2017-00001
- Event Type
- Malfunction
- Date Received
- December 20, 2017
- Date of Event
- December 15, 2017
- Report Date
- December 20, 2017
- Manufacturer
- BK MEDICAL APS
- Product Code
- IYN
- UDI-DI
- 05704916001025
- PMA / PMN Number
- K081154
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- SERVICE PERSONNEL
Narratives
UNIT WAS REPLACED WITH A LOANER UNIT. THE FACILITY WOULD NOT ALLOW THE SERVICE TECHNICIAN TO REMOVE THE DEVICE FROM THEIR PREMISES.
IN THE OR DURING A ROBOTIC SURGERY CASE WITH THE USE OF THE BK MEDICAL ULTRASOUND SYSTEM AT APPROXIMATELY 10:30 AM, WITH THE ULTRASOUND PLUGGED INTO THE WALL, A NURSE NOTICED A NOISE AND SAW SMOKE COMING FROM THE BACK OF FLEX FOCUS 800 ULTRASOUND S/N (B)(4) WITH UA1214 KEYBOARD STAND S/N (B)(4). THE ORDERLY IMMEDIATELY UNPLUGGED THE UNIT FROM THE WALL AND BROUGHT INTO THE HALLWAY. AT THIS TIME, THE UNIT WAS STILL SMOKING AND THE ORDERLY USED A WATER FIRE EXTINGUISHER ON THE ULTRASOUND CONCENTRATING ON THE BATTERY BAY AREA ON THE UA1214. NEXT A FIRE ALARM WAS PULLED AND THE BIO-MED TECH SHOWED UP WITH A REPLACEMENT FLEX 800. THE BIO-MED TECH SAW THE BATTERIES CATCH FIRE WHEN THE WATER WAS APPLIED. THE FIRE DID GO OUT AFTER A SHORT TIME. WHEN THE FIRE DEPT. SHOWED UP THE FIRE WAS OUT. THEY REMOVED THE EQUIPMENT AND BROUGHT IT DOWN TO THE BIO-MED SHOP. THE BIO-MED ENGINEER DID NOT KNOW THE STATUS OF THE CHARGE BEFORE INCIDENT. THEY DID SAY THE NURSES SAW THE WARNING TO REPLACE THE BATTERIES ON THE SCREEN. THE BIOMED DEPARTMENT HAD JUST ORDERED 6 NEW BATTERIES TO REPLACE ON ALL 3 ULTRASOUNDS BUT DID NOT HAVE A CHANCE TO REPLACE ON THIS UNIT YET. THEY WERE NOT AWARE THERE WERE 4 BATTERIES IN THIS UNIT WHEN HE ORDERED THEM. BIO-MED HAD NO KNOWLEDGE OF DAMAGE TO THE BATTERIES OR THE ULTRASOUND BEFORE THE INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 915750 | FLEX FOCUS 800 SYSTEM | ULTRASOUND PULSED DOPPLER IMAGING SYSTEM | IYN | BK MEDICAL APS | FLEX FOCUS 800 SYSTEM | N/A | 05704916001025 |