FDA Adverse Event Injury Summary report: N

M2A-MAGNUM 42-50MM TPR INSRT-3

MDR report key: 7132628 · Received December 19, 2017

Report

Report Number
0001825034-2017-11297
Event Type
Injury
Date Received
December 19, 2017
Date of Event
March 16, 2015
Report Date
December 19, 2017
Manufacturer
ZIMMER BIOMET, INC.
Product Code
KWA
PMA / PMN Number
P042037
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT IS BEING SUBMITTED TO RELAY ON INITIAL INFORMATION AND INVESTIGATION RESULTS. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2017-02006 MEDICAL DEVICES: M2A-MAGNUM PF CUP 52ODX46ID CATALOG#: US157852 LOT: 867630, M2A-MAGNUM MOD HD SZ 46MM CATALOG#: 157446 LOT#: 178590, BI-METRIC/X POR NC 9X125 CATALOG#: X180309 LOT#: 166270. REPORTED EVENT WAS CONFIRMED THROUGH REVIEW OF MEDICAL RECORDS. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN RISK DOCUMENTATION. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY 5 YEAR POST-IMPLANTATION DUE TO PAIN, LACK OF MOBILITY, AND ELEVATED METAL ION LEVELS. DURING THE REVISION, SOME CORROSION WAS FOUND AT THE BASE OF THE HEAD. THERE WAS ALSO SIGNIFICANT CORROSION ON THE TRUNNION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
911078 M2A-MAGNUM 42-50MM TPR INSRT-3 PROSTHESIS, HIP KWA ZIMMER BIOMET, INC. N/A 601660

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| R