FDA Adverse Event Malfunction Summary report: N

CATALYS SYSTEM

MDR report key: 7132219 · Received December 19, 2017

Report

Report Number
3005675890-2017-00096
Event Type
Malfunction
Date Received
December 19, 2017
Date of Event
February 21, 2017
Report Date
January 23, 2018
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
OOE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE RECORDS RELATED TO THE DEVICE THAT INCLUDED LABELING, MANUALS, TRENDING, AND RISK DOCUMENTATION WAS PERFORMED. THE SYSTEM AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. THE TREND REVIEW SHOWS THAT THERE IS NO SIGNIFICANT CHANGE OVER HISTORICAL COMPLAINT LIMITS AND NO RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(6). AMO¿S INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE AMO HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. MANUFACTURING YEAR 2014. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT; LASER CORRECTION SYSTEM, MODEL CATALYS-I, THAT HAS A SIMILAR PRODUCT, CATALYS-U WHICH IS DISTRIBUTED IN THE UNITED STATES UNDER 510(K) # K121091. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED.

Description of Event or Problem · 1

ACCOUNT REPORTED THAT SYSTEM EXPERIENCED SUCTION LOSS WHILE LASER WAS FIRING DURING SURGERY. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
909390 CATALYS SYSTEM CATALYS OOE ABBOTT MEDICAL OPTICS CATALYS-I

Patients

Seq Age Sex Outcome Treatment
1