LMA PROSEAL, REU, SIZE 4 (150040)
Report
- Report Number
- 9681900-2017-00048
- Event Type
- Malfunction
- Date Received
- December 19, 2017
- Date of Event
- December 1, 2017
- Report Date
- December 5, 2017
- Manufacturer
- TELEFLEX MEDICAL
- Product Code
- CAE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION. A VISUAL EXAM WAS PERFORMED AND IT WAS OBSERVED THAT THE SAMPLE WAS DISCOLORED. UPON CLOSER OBSERVATION AT THE FAILURE LOCATION, IT WAS OBSERVED THERE WERE SOME JAGGED EDGES (SYMPTOMS OF CRACKS DUE TO FORCE) ON THE BROKEN CONNECTOR. THE JAGGED EDGES AT WHERE THE CONNECTOR SPLIT APPEARED TO BE RUPTURED BY FORCE. THE DHR WAS REVIEWED AND THERE WERE NO ISSUES FOUND THAT COULD RELATE TO THE REPORTED COMPLAINT. THE CONCENTRATION OF DETERGENT AND HOLDING TIME OF THE AUTOCLAVING SETTINGS WERE UNKNOWN. A CONCENTRATED DETERGENT OR OVERRUN HOLDING TIME COULD HAVE POSSIBLY ADVERSELY IMPACTED THE STURDINESS OF THE CONNECTOR WHEN EXTERNAL FORCE WAS APPLIED. A CONCLUSION CODE COULD NOT BE FOUND AS THE COMPLAINT WAS CONFIRMED; HOWEVER, A ROOT CAUSE COULD NOT BE ESTABLISHED.
(B)(4). THE DEVICE INVOLVED IN THIS COMPLAINT HAS NOT BEEN RECEIVED BY THE MANUFACTURER AT THE TIME OF THIS REPORT. THE INVESTIGATION INTO THIS COMPLAINT IS STILL IN PROGRESS.
CUSTOMER COMPLAINT ALLEGES "THE END OF THE TUBE HAS SNAPPED OFF WHILST IN THE PATIENTS AIRWAY." REQUIRED MEDICAL INTERVENTION WAS TO REPLACE THE LMA. PATIENT CONDITION WAS REPORTED AS "FINE".
CUSTOMER COMPLAINT ALLEGES "THE END OF THE TUBE HAS SNAPPED OFF WHILST IN THE PATIENTS AIRWAY." REQUIRED MEDICAL INTERVENTION WAS TO REPLACE THE LMA. PATIENT CONDITION WAS REPORTED AS "FINE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 908216 | LMA PROSEAL, REU, SIZE 4 (150040) | AIRWAY,OROPHARYNGEAL,ANESTHESIOLOGY | CAE | TELEFLEX MEDICAL | 8VRA2Q56 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |