FDA Adverse Event
Malfunction
Summary report: N
NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM
MDR report key: 713050
·
Received November 22, 2005
Report
- Report Number
- 6000034-2005-00337
- Event Type
- Malfunction
- Date Received
- November 22, 2005
- Date of Event
- August 24, 2004
- Report Date
- November 22, 2005
- Manufacturer
- COCHLEAR LTD.
- Product Code
- MCM
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
IN SEPTEMBER 2004, THE PT WAS IN AN AUTOMOBILE ACCIDENT. FOLLOWING THE ACCIDENT, SHE REPORTED PAIN AND DECREASED SOUND. QUALITY. AN INTEGRITY TEST SHOWED UNSUAL RESULTS ON 3 ELECTRODES. A SPEECH PROCESSOR PROGRAM WITH THESE ELECTRODES DELATED RESOLVED THE COMPLAINT OF PAIN. A SECOND INTEGRITY TEST WAS DONE 08/2005. IT WAS DETERMINED THAT MULTIPLE ELECTRODES WERE NOT FUNCTIONING APPROPRIATELY
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM | COCHLEAR IMPLANT | MCM | COCHLEAR LTD. | CI24R(CS) | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Required Intervention |