COBAS 8000 E 602 MODULE
Report
- Report Number
- 1823260-2017-02987
- Event Type
- Malfunction
- Date Received
- December 19, 2017
- Date of Event
- November 21, 2017
- Report Date
- December 27, 2017
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K100853
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE EXPIRATION DATE FOR THE CA 19-9 REAGENT LOT IN USE WAS OCT-2018. A SPECIFIC ROOT CAUSE WAS NOT IDENTIFIED. CALIBRATION AND QC WERE ACCEPTABLE. SEVERAL "ABNORMAL SAMPLE ASPIRATION" MESSAGES FROM THE DAY OF THE EVENT WERE OBSERVED DURING A REVIEW OF THE ALARM TRACE. THE CUSTOMER USED A 13X75 SAMPLE TUBE WITHOUT A RACK ADAPTER. POSSIBLE ROOT CAUSES FOR THE ISSUE MAY BE RELATED TO THE CUSTOMER NOT USING A RACK ADAPTER FOR THE 13MM SAMPLE TUBE OR BUBBLES OR FOAM ON THE SAMPLE SURFACE. ADDITIONAL POSSIBLE ROOT CAUSES FOR THIS EVENT MAY BE SAMPLE QUALITY AND INSUFFICIENT MAINTENANCE. A GENERAL PRODUCT PROBLEM HAS BEEN EXCLUDED.
(B)(6). (B)(4).
THE CUSTOMER COMPLAINED OF AN ERRONEOUS LOW RESULT FOR 1 PATIENT TESTED FOR ELECSYS CA 19-9 IMMUNOASSAY (CA 19-9) ON A COBAS 8000 E 602 MODULE. THE INITIAL CA 19-9 RESULT FROM AN ALIQUOT TUBE WAS 14.6 KU/L. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. ON (B)(6) 2017 THE CUSTOMER REPEATED THE SAMPLE ON THE SAME E602 MODULE AND THE RESULT WAS 5678 KU/L. THERE WAS NO ALLEGATION THAT AN ADVERSE EVENT OCCURRED. THE CA 19-9 REAGENT LOT NUMBER WAS 231612. THE EXPIRATION DATE WAS NOT PROVIDED. THE CUSTOMER STATED ALL INSTRUMENT MAINTENANCE IS UP TO DATE. LIQUID FLOW CLEANING IS PERFORMED WEEKLY AND PINCH TUBING WAS CHANGED ON (B)(6) 2017.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 906938 | COBAS 8000 E 602 MODULE | IMMUNOCHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | E602 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |