FDA Adverse Event Malfunction Summary report: N

HELICAL BLADE INSERTER

MDR report key: 7126563 · Received December 18, 2017

Report

Report Number
2939274-2017-50354
Event Type
Malfunction
Date Received
December 18, 2017
Date of Event
November 20, 2017
Report Date
November 20, 2017
Manufacturer
WRIGHTS LANE : SYNTHES USA PRODUCTS LLC
Product Code
LXH
UDI-DI
10886982196064
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SERVICE AND REPAIR EVALUATION WAS COMPLETED. THE CUSTOMER REPORTED THE INSTRUMENT BECAME WARPED. THE REPAIR TECHNICIAN REPORTED THE ALIGNMENT INDICATOR WAS MISSING, AND THE HANDLE WAS SCARRED. MISSING PARTS IS THE REASON FOR REPAIR. THE CAUSE OF THE ISSUE IS UNKNOWN. THE FOLLOWING PARTS WERE REPLACED: REPLACEMENT ALIGNMENT. THE ITEM WAS REPAIRED PER THE INSPECTION SHEET, PASSED SYNTHES FINAL INSPECTION ON 2-JAN-2018 AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. THE EVALUATION WAS CONFIRMED. THE PREVIOUS SERVICE EVENT FOR PART NUMBER 357.372 WITH LOT NUMBER(S) 5052982 HAS BEEN REVIEWED. THE CUSTOMER CALLED IN A SERVICE REQUEST FOR THIS ITEM ON (B)(6) 2017 FOR INOPERABLE - WARPED. THE ITEM WAS PREVIOUSLY RETURNED FOR SERVICE ON (B)(6) 2017 DUE TO ETCH UNREADABLE. THE PREVIOUS SERVICE CONDITION OF ETCH UNREADABLE IS NOT RELEVANT TO THE CURRENT COMPLAINED ISSUE OF INOPERABLE - WARPED. THE MANUFACTURE DATE OF THIS ITEM IS 29-JUL-2005. THE SERVICE HISTORY REVIEW IS UNCONFIRMED. DEVICE HISTORY RECORDS REVIEW WAS COMPLETED FOR PART # 357.372, LOT # 5052982. MANUFACTURING LOCATION: (B)(4), RELEASE TO WAREHOUSE DATE: JUL 29, 2005. NO NON-CONFORMANCE REPORTS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. PATIENT WEIGHT AND DATE OF BIRTH ARE NOT AVAILABLE FOR REPORTING. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. WITHOUT A LOT NUMBER THE DEVICE AND SERVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED A PATIENT HAD INITIAL SURGERY ON (B)(6) 2017 FOR TREATMENT OF A SUB-TROCHANTERIC FEMUR FRACTURE. PATIENT WAS IMPLANTED WITH ONE (1) TROCHANTERIC FIXATION NAIL (TFN), ONE (1) 11.0MM TI HELICAL BLADE AND ONE (1) 4.5MM DISTAL LOCKING SCREW. AS THE SURGEON WAS USING THE SLIDE HAMMER WHILE INSERTING THE HELICAL BLADE IMPLANT WITH THE HELICAL BLADE COUPLING SCREW, THE DISTAL KNOB ON THE COUPLING SCREW BROKE OFF INTRAOPERATIVELY, AND FELL ON THE OPERATING ROOM FLOOR. DUE TO THE FORCE THAT THE SURGEON USED TO DISENGAGE THE HELICAL BLADE IMPLANT FROM THE TFN INSTRUMENTS, THE AIMING ARM INSTRUMENT AND THE HELICAL BLADE INSERTER INSTRUMENT BECAME WARPED. ALSO, THE BUTTRESS COMPRESSION NUT BECAME STUCK INSIDE THE BLADE GUIDE SLEEVE. THE SURGEON SUCCESSFULLY DISENGAGED ALL THE INSTRUMENTS FROM THE IMPLANT, AND PROCEEDED WITH THE SURGERY. NO FRAGMENTS WERE GENERATED. SURGERY WAS COMPLETED SUCCESSFULLY WITH 30-40 MINUTE TIME DELAY. PATIENT IS REPORTED IN STABLE CONDITION. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
903800 HELICAL BLADE INSERTER MISC ORTHO SURGICAL INSTR LXH WRIGHTS LANE : SYNTHES USA PRODUCTS LLC 5052982 10886982196064

Patients

Seq Age Sex Outcome Treatment
1 69 YR