VITEK® 2 AST-YS08 TEST KIT.
Report
- Report Number
- 1950204-2017-00461
- Event Type
- Malfunction
- Date Received
- December 18, 2017
- Report Date
- May 18, 2018
- Manufacturer
- BIOMERIEUX, INC
- Product Code
- NGZ
- PMA / PMN Number
- K061945
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A CUSTOMER FROM THE (B)(6) REPORTED TO BIOMÉRIEUX AN OUT OF RANGE LOW (OORL) FLUCONAZOLE (FLU) RESULT FOR A CANDIDA KRUSEI QUALITY CONTROL STRAIN (ATCC 62058), IN ASSOCIATION WITH THE VITEK® 2 AST-YS08 TEST KIT. THE CUSTOMER SUBMITTED THE STRAIN AND CARDS FOR EVALUATION. AN INVESTIGATION WAS PERFORMED. ON VTEK 2 (V7.01), TESTS WERE PERFORMED WITH TWO C. KRUSEI ATCC 6258 STRAINS (INTERNAL REFERENCE STRAIN AND CUSTOMER STRAIN), FROM SDA SUBCULTURED UNDER AEROBIC ATMOSPHERE AS RECOMMENDED. TWO (2) AST-YS08 CARDS LOTS WHICH INCLUDED THE CUSTOMER LOT (2880150203) AND RANDOM LOT (2880297203) WERE TESTED IN DUPLICATE. THE FLU RESULTS WERE MIC = 8 MG/L FOR BOTH LOTS AND WITH BOTH STRAINS TESTED. THE CUSTOMER ISSUE WAS NOT DUPLICATED. THE VITEK 2 AST-YS08 CARD PERFORMED AS INTENDED AND NO FURTHER ACTION IS REQUIRED.
A CUSTOMER FROM THE (B)(6) REPORTED TO BIOMÉRIEUX AN OUT OF RANGE LOW (OORL) FLUCONAZOLE RESULT FOR A CANDIDA KRUSEI QUALITY CONTROL STRAIN (ATCC 62058), IN ASSOCIATION WITH THE VITEK® 2 AST-YS08 TEST KIT. THE CUSTOMER TESTED THE STRAIN TWICE, BY USING THE AST-YS08 CARD ON TWO DIFFERENT INSTRUMENTS. THE RESULT WAS MIC = 4 FOR BOTH TESTS WITH THE EXPECTED RANGE OF MIC 8 >/= 64. THERE WAS NO PATIENT INVOLVEMENT AS THE EVENT PERTAINED TO A QUALITY CONTROL SAMPLE. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 904207 | VITEK® 2 AST-YS08 TEST KIT. | VITEK® 2 AST-YS08 TEST KIT. | NGZ | BIOMERIEUX, INC | 2880150203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |