FDA Adverse Event Malfunction Summary report: N

VITEK® 2 AST-YS08 TEST KIT

MDR report key: 7120367 · Received December 15, 2017

Report

Report Number
1950204-2017-00454
Event Type
Malfunction
Date Received
December 15, 2017
Report Date
May 11, 2018
Manufacturer
BIOMERIEUX INC.
Product Code
NGZ
PMA / PMN Number
K133952
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT WAS INITIALLY SUBMITTED IN RESPONSE TO A CUSTOMER IN (B)(6) REPORTING FALSE SUSCEPTIBLE FLUCYTOSINE (ALSO CALLED FLUOROCYTOSINE) RESULTS FOR A CANDIDA NORVEGENSIS EXTERNAL QUALITY CONTROL (EEQ) STRAIN IN ASSOCIATION WITH THE VITEK® 2 AST-YS08 TEST KIT. THE RESULT WAS FLUCYTOSINE : MIC = 4 SUSCEPTIBLE. THE EXPECTED RESULT WAS RESISTANT. AN INTERNAL BIOMÉRIEUX INVESTIGATION WAS PERFORMED THAT INCLUDED TESTING FOR BOTH FLUCONAZOLE (FLU) AND FLUCYTOSINE (FCT) RESULTS WITH A STRAIN OF CANDIDA NORVEGENSIS ON VITEK® 2 AST-YS08 CARDS. THIS STRAIN IS FROM A BIOLOGIE PROSPECTIVE SURVEY (MYCOLOGY 2017- SAMPLE 3A). ID TESTING : IDENTIFICATION WAS CONFIRMED TO CANDIDA NORVEGENSIS ON VITEK 2 YST ID CARD (LOT 2430356103). AST TESTING: · REFERENCE METHOD TO DETERMINE THE INTENDED RESULT FOR FLUCONAZOLE AND FLUCYTOSINE: BROTH MICRODILUTION, METHOD USED TO DEVELOP THE FORMULARIES "FLU02N" / "FCT02N" (IN AST-YS08 CARD). · FLU MIC = 16 MG/L INTERMEDIATE · FCT MIC = 8 MG/L INTERMEDIATE INTERPRETATION ACCORDING TO VITEK 2 BREAKPOINTS IN V7.01 : FDA 2009 BREAKPOINTS FOR CANDIDA / FLUCONAZOLE S </=8 - I 16-32 - R >/=64. FDA 2012 BREAKPOINTS FOR CANDIDA / FLUCYTOSINE : S </=4 - I 8-16 - R >/=32. NOTE: FOR ANTIFUNGALS, ESSENTIAL AGREEMENT BETWEEN VITEK® 2 CARD AND REFERENCE METHOD MICS IS +/- 2 DOUBLING DILUTIONS. · PRODUCTS & SYSTEM INCRIMINATED: -AST -YS08 ON VITEK 2 V7.01 THREE (3) VITEK 2 AST-YS08 LOTS WERE TESTED FROM SABOURAUD DEXTROSE AGAR (SDA). VITEK 2 GAVE THE FOLLOWING RESULTS ON THE THREE (3) LOTS TESTED: **FLU MIC = 8 MG/L SUSCEPTIBLE **FCT MIC = 4 MG/L SUSCEPTIBLE THE VITEK 2 RESULTS ARE WITHIN ESSENTIAL AGREEMENT WITH THE REFERENCE MIC WITHIN +/- 1 DOUBLING DILUTION. CONCLUSION: -THE CUSTOMER RESULTS WERE DUPLICATED ON THE AST-YS08 CARDS (MIC FLU= 8 MG/L S AND FCT MIC=4 MG/L S). -THE VITEK 2 AST-YS08 CARDS PERFORM AS INTENDED. THE VITEK 2 VALUES ARE WITHIN ESSENTIAL AGREEMENT WITH THE REFERENCE RESULTS OBTAINED IN BMD. -THE REFERENCE MIC FOR BOTH FLUCONAZOLE AND FLUCYTOSINE ARE IN THE MIDDLE OF THE BREAKPOINTS, SO THE STRAIN CAN BE SUSCEPTIBLE OR RESISTANT WITHIN +/- 2 DOUBLING DILUTIONS. -->BORDERLINE STRAIN.

Description of Event or Problem · 1

A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF A FALSE SUSCEPTIBLE RESULT WHILE TESTING AN EXTERNAL QUALITY CONTROL (EQC) (REF. 2017-3) FROM BIOLOGICAL PROSPECTIVE FOR ANTIFUNGAL SUSCEPTIBILITY USING BIOMERIEUX VITEK®2 AST-YS08. THE CUSTOMER TESTED THE EQC SAMPLE FOR IDENTIFICATION USING VITEK®2 AST-YS08 CARD AND CANDIDA NORVEGENSIS WAS IDENTIFIED. THE CUSTOMER REPORTED THAT THEY HAD TESTED AND SUBCULTURED THE STRAIN AND OBTAINED: ·FLUOROCYTOSINE: MIC VALUE OF 4 INDICATING THE FUNGUS TO BE SUSCEPTIBLE TO FLUOROCYTOSINE WHEREAS THE EXPECTED RESULT WAS RESISTANT. ·FLUCONAZOLE: MIC VALUE OF 16 INDICATING INTERMEDIATE WHEREAS THE EXPECTED RESULT WAS RESISTANT. THE MEDIUM OF CULTURE WAS SABOURAUD GENTAMICINE CHLORAMPHÉNICOL 2. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. THERE WAS NO PATIENT ASSOCIATED WITH THIS SURVEY ISOLATE. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
902441 VITEK® 2 AST-YS08 TEST KIT VITEK® 2 AST-YS08 TEST KIT NGZ BIOMERIEUX INC. 2880150203

Patients

Seq Age Sex Outcome Treatment
1