PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAH
Report
- Report Number
- 2032227-2017-73505
- Event Type
- Malfunction
- Date Received
- December 14, 2017
- Date of Event
- August 9, 2017
- Report Date
- January 2, 2018
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169513884
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PATIENT
Narratives
FINDINGS: THE INSULIN PUMP HAD BLANK DISPLAY DUE TO MOISTURE DAMAGE ON LCD BOARD. UNABLE TO PERFORM IDLE CURRENT TEST, RUN CURRENT TEST, SELF TEST, OFF NO POWER TEST, UNEXPECTED RESTART ERROR TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/COMPROMISED FORCE SENSOR TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT TEST OR VERIFY BATTERY OUT LIMIT ALARM DUE TO BLANK DISPLAY. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP. (B)(4).
PUMP HAD BLANK DISPLAY DUE TO MOISTURE DAMAGE ON LCD BOARD. UNABLE TO PERFORM IDLE CURRENT TEST, RUN CURRENT TEST, SELF TEST, OFF NO POWER TEST, UNEXPECTED ALARM ERROR TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIMETEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT TEST OR VERIFY BATTERY OUT LIMIT ALARM DUE TO BLANK DISPLAY. PUMP RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.
CUSTOMER REPORTED VIA PHONE CALL THAT THE DEVICE HAD ALARMED BATTERY OUT LIMIT ALARM AND HAD A BLANK DISPLAY. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONALS¿ INSTRUCTIONS. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. CUSTOMER AGREED TO RETURN THE DEVICE FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 895678 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAH | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAH | A5723LNAHJ | 00643169513884 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |