T:SLIM G5 SYSTEM
Report
- Report Number
- 3007981285-2017-38964
- Event Type
- Malfunction
- Date Received
- December 14, 2017
- Date of Event
- November 30, 2017
- Report Date
- December 14, 2017
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OYC
- PMA / PMN Number
- P140015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PATIENT
Narratives
T:SLIM X2 USER GUIDE STATES, "ALWAYS MAKE SURE THAT THE CORRECT TIME AND DATE ARE SET ON YOUR PUMP. WHEN EDITING TIME, ALWAYS CHECK THAT THE AM/PM SETTING IS ACCURATE. NOT HAVING THE CORRECT TIME AND DATE SETTING MAY AFFECT SAFE INSULIN DELIVERY." THE T:SLIM X2 USER GUIDE STATES, ¿THE SINGLE-USE DISPOSABLE CARTRIDGE CAN HOLD UP TO 300 UNITS (3.0 ML) OF INSULIN¿. THE T:SLIM X2 USER GUIDE STATES, "REPLACE THE CARTRIDGE EVERY 48 HOURS IF USING HUMALOG". NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED BLOOD GLUCOSE (BG) LEVELS IN THE 211-215 MG/DL RANGE. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, IT WAS NOTED THAT THERE WAS AN AIR BUBBLE PRESENT IN THE INFUSION SET TUBING. THE CUSTOMER USES HUMALOG INSULIN, AND REPORTED THE CARTRIDGE AND INFUSION SET ARE TYPICALLY IN USE FOR APPROXIMATELY 5 DAYS AT A TIME. IN ADDITION, THE CUSTOMER PREVIOUSLY INTENTIONALLY OVERFILLED THE CARTRIDGE WITH APPROXIMATELY 320 UNITS OF INSULIN. DURING TROUBLESHOOTING, IT WAS ALSO NOTED THAT THE PUMP TIME WAS INACCURATE, AND THE CUSTOMER DID NOT HAVE TO CORRECT SETTINGS PROFILE TURNED ON. CUSTOMER REPORTED THE TIME WAS INACCURATE DUE TO USE ERROR. A NEW CARTRIDGE WAS LOADED ONTO THE PUMP WITH THE PROPER LOADING PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 898890 | T:SLIM G5 SYSTEM | CONTINUOUS GLUCOSE MONITOR | OYC | TANDEM DIABETES CARE | 1000096 | M020465 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR |